Equipment Description
29
Haemonetics
®
ACP-215
®
Service Manual
P/N: SM-ACP215-01-EN(AB)
Shaker
A dedicated shaker properly mixes the blood in the RBC bag with the added
solutions. The shaker connects to the rear panel of the device through a hard-
wired cable with a 15-pin subminiature “D” connector (see
). The function of each pin is defined in
Figure 8, Pin out for shaker cable (end view)
The device provides power to the shaker and controls the shaker’s operations.
The shaker has a platform to firmly hold the RBC bag in place using magnets.
The shaker operates at a frequency of 3 Hz and 1.5 inch amplitude
displacement, (peak-to-peak).
Note: The shaker unit is not repairable in the field. For technical assistance or
to have your shaker serviced in the USA, call Haemonetics Customer Care
Service at 1-800-537-2802. For technical assistance or to have your shaker
serviced in other countries, please contact your local Haemonetics
representative.
Table 1
Pin Number
Pin Definition
1, 2, 3, 4
Motor
-
output
5, 6, 7, 8
Motor
+
output
9, 10, 11
No connection
12
+5 volts DC
13
DC ground
14
Motor encoder
A
input
15
Motor Encoder
B
input
Содержание ACP 215
Страница 10: ......
Страница 38: ......
Страница 58: ......
Страница 104: ...104 Preventive Maintenance and Installation P N SM ACP215 01 EN AB Haemonetics ACP 215 Service Manual...
Страница 126: ......
Страница 150: ......
Страница 182: ...P N SM ACP215 01 EN AB Haemonetics ACP 215 Service Manual 182 System Schematics System Schematic...
Страница 183: ...Haemonetics ACP 215 Service Manual P N SM ACP215 01 EN AB System Schematics 183 Family Tree...
Страница 184: ...P N SM ACP215 01 EN AB Haemonetics ACP 215 Service Manual 184 System Schematics Pneumatic Schematic...
Страница 186: ......