Equipment Description
27
Haemonetics
®
ACP-215
®
Service Manual
P/N: SM-ACP215-01-EN(AB)
Centrifuge
Figure 6, Centrifuge
The device uses a sophisticated centrifuge assembly that offers key
operational features. The centrifuge cover halves are gear coupled and hinged
to provide exceptional access to the centrifuge well. The cover halves also
secure the bowl header in place. An optical switch reports on the position of the
cover latch mechanism. The fluid sensor used to detect moisture in the
centrifuge well area is mounted in the centrifuge well (see
Excellent drainage is provided in the event of a fluid spill, which terminates in
a disposable waste bag located at the rear of the device.
The centrifuge motor is controlled by a centrifuge motor control PCB,
discussed later in this chapter.
Circuit Description
One optical sensor controls a +15 VDC signal and the other optical sensor
controls a +28 VDC signal. The +28 VDC is used by the centrifuge distribution
card to energize a relay, which directly controls the +48 VDC centrifuge drive
source. Both signals are then optically isolated and monitored on the CPU card
through an I/O port.
Содержание ACP 215
Страница 10: ......
Страница 38: ......
Страница 58: ......
Страница 104: ...104 Preventive Maintenance and Installation P N SM ACP215 01 EN AB Haemonetics ACP 215 Service Manual...
Страница 126: ......
Страница 150: ......
Страница 182: ...P N SM ACP215 01 EN AB Haemonetics ACP 215 Service Manual 182 System Schematics System Schematic...
Страница 183: ...Haemonetics ACP 215 Service Manual P N SM ACP215 01 EN AB System Schematics 183 Family Tree...
Страница 184: ...P N SM ACP215 01 EN AB Haemonetics ACP 215 Service Manual 184 System Schematics Pneumatic Schematic...
Страница 186: ......