Running the 2823 first time
Performing a manual measurement
DUT Data introduction
DUT data
1. Go to Setup -> DUT Info and Introduce the desired
information, no field is mandatory.
Tan
δ
(power factor) varies with the temperature for some
test objects. The 2823 can automatically compensate this
variation and refers the values at reference temperature.
1. In Ambient->Temperature introduce the ambient
temperature
2. In Device under test DUT insulation->temperature
introduce the reference temperature
3. Select the test object type
4. The temperature correction table will arise in the
button
Measuring
DUT data
1. Go to Manual
2. Select the parameter to be shown on each
indicator by clicking in the small triangle after the
name
1
2
3
4
1
2