21
Technical Data
0775 Veloria MD
total weight: 2,7 kg
dimension: W273xH142xD186 mm
input: 100 V - 240 V~, 50/60 Hz
power consumpt.: 1A(100V)-0,5A(240V)
fuse primary: F1, F2: 4 A time delay
fuse sec.: thermal fuse
protection class I
accuracy of speed display: ±10 %
handpiece:
type B applied part
diameter: 19 - 25 mm
length: 144 mm
speed: 6.000 - 40.000 rpm
Ambient Conditions
The device is intended for use in
dry closed rooms.
Operation:
Temp.: +10°C to +26°C
Rel. humidity: 25% to 75%
(not condensed)
Air pressure: 700 hPa to 1060 hPa
Storage:
Temp: -5°C to +55°C
Rel. humidity: 10% to 95%
(not condensed)
Air pressure: 500 hPa to 1060 hPa
Transport (up to 4 weeks):
Temp: -5°C to +55°C
Rel. humidity: 10% to 95%
(not condensed)
Air pressure: 500 hPa to 1060 hPa
Operating Time
The device is designed for extend-
ed usage.
Accessories & Spare Parts
art.
5179
micro filter
3752
control unit filter
0970
foot switch
6158
cleaning brush
The dust filter (art. 5179)
must be used. The paper
filter (art. 5115) must NOT
be used.
The clamping system is designed
for rotating instruments, which have
a shaft of 2,35 mm in accordance
with DIN EN ISO 1797-1.
Compulsory Registration
Operators, distributors and retailers
who were informed about the occur-
rences regarding §29 of the Medical
Devices law must report this. The
method of reporting is detailed in
the Medical Devices Safety Plan
Ordinance. Outside of Germany the
corresponding local conditions ap-
ply.