14
Application
Exchange control unit filter
This filter (Art. No. 3752) has to be
exchanged after every 200 operat-
ing hours. Remove the old filter and
place the new one in its holder (see
picture).
Clamping the tool/ bur
The handpiece is equipped with a
chuck, which can hold all burs with
a diameter of 2.35 mm (this is the
diameter usually used for tools in
podiatry). To open the chuck, push
the knob forward with your thumb
and leave it there. Now insert the
tool as far as possible. Pull back the
knob and the tool will be firmly
clamped
For security reasons, only
change the tool, if the hand-
piece is turned off.
Setting up the speed
Hold the handpiece (incl. tool) in
your hand or store it in the hand-
piece holder attached to the side of
the control unit.
Switch on the unit with button (9).
Now you can also set the speed
with buttons (1) and (2), and choose
its direction of rotation with button
(6).
When the device is brought
out of standby mode it will
run with the rotation speed
and suction power used last. Take
care to not start the mill at an overly
high rotation speed. Select the re-
quired rotation speed before chang-
ing the milling tool.
Footswitch (0970)
Operation of the footswitch:
Only use the original hadewe foot
switch, to avoid any damages.
Insert the plug of the foot switch into
the socket with the symbol
.
If you press the foot switch for a
longer period, you will switch on /
off the complete control unit.