15
Application
Infilling the spray liquid
Switch off the device. Pull out the
spray tank on its handle. Hold the
cover of the tank in hand and un-
screw the tank.
Now fill it with distilled water and
anti germ liquid (e.g. AQUAHY item
no. 8555).
Do not exceed the maxi-
mum level when filling
the tank.
Now turn the tank cover back on the
tank. While screwing back the cover
on the tank, make sure that is
closed correct. Otherwise there will
no pressure be build up.
Put the tank back in the housing.
Switch on the power switch in the
rear of the unit.
Attention:
Never use tap
water. (Danger of calcifi-
cation). Also do not use
explosive or combustible liquids e.g.
alcohol.
Note:
The filter in the tank (item no.
3479) should be fully under pres-
sure; otherwise no water will be
sucked in.
The water filter has to be
fully under water, otherwise
no water will be sucked in.
Mounting/Changing the
filter
Remove cover plate with filter
(Art.5115). Now the filter can be
pulled from the plate and be
changed. Never work with full filter
or without filter. Put up cover plate.
The cover plate must lie flat but
may move a little. It fits tightly when
the suction cleaner is in operation.
As an option you may also use the
micro filter (Art. No. 5179) instead
of paper filter 5115.
Insert the bur
The handpiece is equipped with a
chuck which can hold all burs with a
diameter of 2.35 mm and usually
are used in the fields of podiatry. To
open the chuck, push the black
knob forward with your thumb and
leave it there. Now insert the tool as
far as possible. Pull back the black
knob and the tool will be clamped.