Hach TU5200 Скачать руководство пользователя страница 35

Message

Solution

The sample ID list full. Data
has not been added

Remove the sample IDs that are not used. Add a new sample ID.

Documents/Videos

The document does not
show on the display.

Make sure that the document is saved as a PDF.

Not enough memory to store
the document.

Delete other documents, then add the document again.

Cannot delete the
document.

Contact technical support.

Cannot store the document.

Contact technical support.

Cannot open the manual.

Make sure that the document format is PDF.

The video format is not
supported.

Make sure that the video format is AVI (Codec XVID) with a resolution of 640 x
480 pixels. Make sure that the video size is not more than 10 Mb.

The video does not play.

The video file is corrupt.

Ethernet/LAN connection

Network authentication
failed

Enter the correct username and password. Make sure that the network settings
are correctly configured.

LAN Connection failed

Automatic retrieval of the
local IP address failed!

There is no connection to the DHCP server. Contact the network administrator.
If the DHCP server is not enabled, manually enter an IP address for the
instrument.

Setting IP-address failed!

Contact the network administrator. If necessary, contact technical support.

Setting subnet-mask failed!

Setting default gateway
failed!

Mounting network drive
failed!

FTP connection failed!

Verify network configuration

Files did not save to the
network share.

Make sure that the correct permissions are set to the network share directory.

No connection to the web
server.

Try to connect the instrument to a LAN again later. Make sure that the instrument
is connected to the network.

RFID

Standard expired. Continue? Use standards that are not expired.

Primary standard is write
protected.

The verification standard value cannot be recorded on a calibration standard.

Measured value is out of
nominal value range.

The measured value is not within range for this standard. The 10 NTU standard
range is 5 to 15 NTU. The < 0.1 NTU glass verification rod range is 0 to 0.1 NTU.
Use the correct standard. Make sure that the standard is not expired. Measure the
standard again.

No act. value on RFID tag.
Read standard?

Measure the verification standard first on an instrument that was recently
calibrated.

English

   

33

Содержание TU5200

Страница 1: ...DOC342 53 80489 TU5200 03 2017 Edition 3 User Manual...

Страница 2: ......

Страница 3: ...perator IDs 13 Configure an operator RFID tag optional 14 Add sample IDs 14 Import sample IDs optional 14 Configure the measurement settings 15 Set the acceptance range 15 Set up a LAN connection 16 C...

Страница 4: ...gs 27 Do a calibration verification 28 Maintenance 28 Clean spills 29 Clean the instrument 29 Clean a sample vial 29 Clean the vial compartment 30 Troubleshooting 30 Remove air bubbles from the sample...

Страница 5: ...EBC Accuracy 2 of reading plus 0 01 NTU from 0 to 40 FNU 10 of reading from 40 to 1000 FNU based on Formazin primary standard at 25 C 77 F Linearity Better than 1 for 0 to 40 NTU on Formazin at 25 C...

Страница 6: ...ims such damages to the full extent permitted under applicable law The user is solely responsible to identify critical application risks and install appropriate mechanisms to protect processes during...

Страница 7: ...electrocution exists This symbol indicates the need for protective eye wear This symbol indicates a laser device is used in the equipment This symbol identifies a risk of chemical harm and indicates...

Страница 8: ...se of doubt contact the manufacturer Safety information for RFID modules W A R N I N G Multiple hazards Do not disassemble the instrument for maintenance If the internal components must be cleaned or...

Страница 9: ...ncy energy and if not installed and used in accordance with the instruction manual may cause harmful interference to radio communications Operation of this equipment in a residential area is likely to...

Страница 10: ...rt for LAN connection 4 Power button 9 RFID module indicator optional 5 Power supply connection 10 USB port type A Product components Make sure that all components have been received Refer to Figure 2...

Страница 11: ...lation to break down which can result in an electric shock hazard The manufacturer recommends that users with concerns contact technical support Installation guidelines Install the instrument On a lev...

Страница 12: ...o not use writing tips of pens or pencils or other sharp objects to make selections on the screen or damage to the screen will occur Refer to Figure 3 for an overview of the home screen Figure 3 Displ...

Страница 13: ...ading log calibration log verification log and compare log Refer to Show the recorded data on page 24 Setup Configures the instrument settings Refer to Configure the instrument settings on page 12 Dia...

Страница 14: ...mmm yyyy default yyyy mm dd dd mm yyyy or mm dd yyyy Time Format Sets the time format Options 12 or 24 hours default Security Enables or disables password protection for the settings and tasks in the...

Страница 15: ...nguage and then push OK Add operator IDs Add a unique operator ID for each person who will measure samples 30 maximum Select an icon operator password and security level for each operator ID 1 Push Lo...

Страница 16: ...e barcode with a barcode handset scanner that is connected to the instrument The barcode is added to the sample ID 5 Push OK 6 Select an option Option Description Add Date Time Adds the data and time...

Страница 17: ...s the measurement units that show on the display and that are recorded to the data log Options NTU FNU TE F FTU EBC mNTU or mFNU Default FNU Data Log Setup Sets the data log settings Auto Store Measur...

Страница 18: ...Select Network then push Setup 5 Select On 6 Push TU5200 IP 7 To get an IP address for the instrument from a DHCP server do the steps that follow a Select Automatic b Push OK two times The IP address...

Страница 19: ...network printers that are SNMP enabled and in the same sub network are found by the instrument 1 Push two times then push Setup 2 Push Network Peripherals 3 Select Printer then push Setup 4 Push Netwo...

Страница 20: ...th the sample When collecting a sample from a water tap in a distribution system or treatment plant turn the water on for at least five minutes then collect the sample Do not adjust the flow because t...

Страница 21: ...l and national regulations N O T I C E Always put a cap on the sample vial to prevent spills in the vial compartment Refer to the illustrated steps that follow to prepare a sample vial for measurement...

Страница 22: ...operator RFID tag in front of the RFID module or Push Login Select the applicable operator ID then push Select 2 Select the sample ID as follows Put the sample RFID sticker on the sample bottle in fro...

Страница 23: ...cess and lab instrument are calibrated with the same number of calibration points and with the same standards Make sure that the calibrations are not expired Collect a grab sample Collect a 100 mL sam...

Страница 24: ...eading from the process instrument shows on the display 11 Prepare a grab sample vial Refer to Prepare a sample vial on page 19 12 Measure the turbidity of the grab sample with the laboratory instrume...

Страница 25: ...e laboratory instrument 7 Move the grab sample bottle to the laboratory instrument 8 On the TU5200 push Options Reading Setup 9 Push Bubble Reject then set bubble reject to on 10 If the grab sample is...

Страница 26: ...Calibration log Shows the calibration history Verification log Shows the verification history Compare log Shows the recorded comparisons of process and laboratory measurements 1 Push Data Log and sel...

Страница 27: ...data point Data points Selects the symbol used for the data points Control Limit Sets the minimum value and maximum value of the readings that show on the graph Show the instrument information 1 Push...

Страница 28: ...er with kaolin SDVB 0 1000 FNU 3 point calibration 20 FNU 600 FNU and dilution water with spherical slyrene divinylbenzene Custom Calibration 1 to 6 point calibration 0 02 to 1000 FNU with StablCal or...

Страница 29: ...measured 12 Push Store to complete the calibration and record the results to the calibration log 13 To show the calibration log push Options Calibration Log Refer to Show the recorded data on page 24...

Страница 30: ...ect 2 Push Verification 3 Clean the verification standard vial with a no lint cloth to remove contamination Keep contamination off the glass Refer to Prevent vial contamination on page 18 4 Dry the ex...

Страница 31: ...current safety data sheets MSDS SDS for safety protocols Clean the sample vial when there is contamination in the sample vial after the sample vial is rinsed Items to collect Hydrochloric acid concen...

Страница 32: ...micro fiber cloth lint free cloth Liquid oil Cloth water and cleaning agent Troubleshooting Message Solution Startup The self check stopped Clean the vial compartment The vial compartment is dirty Set...

Страница 33: ...stable Use the correct calibration standards Invert the standard until no bubbles or large particles show The standard value is out of the measurement range Use the correct calibration standards Inve...

Страница 34: ...that the file system FAT32 is installed on the USB flash drive Set the power to off wait 20 seconds and then set the power to on again Look for remaining space on the USB flash drive Set the power to...

Страница 35: ...k administrator If the DHCP server is not enabled manually enter an IP address for the instrument Setting IP address failed Contact the network administrator If necessary contact technical support Set...

Страница 36: ...ollects on the sample vial during measurement let the sample become slightly warmer before measurement Keep the sample at room temperature or partially put the sample vial into a warm water bath for a...

Страница 37: ...60 Hz 1 LZV798 Sample vials with caps 2 LZV946 Vial stand 1 LZV952 Accessories Description Quantity Item no Micro fiber cloth vial cleaning 1 LZY945 RFID tags operator 2 pkg LZQ066 RFID stickers samp...

Страница 38: ...36 English...

Страница 39: ......

Страница 40: ...orders hach com www hach com HACH LANGE GMBH Willst tterstra e 11 D 40549 D sseldorf Germany Tel 49 0 2 11 52 88 320 Fax 49 0 2 11 52 88 210 info de hach com www de hach com HACH LANGE S rl 6 route de...

Отзывы: