Loosen the tube connector on the bottle cap.
Put a filled desiccant cartridge into the adapter on the titrant and solvent bottle caps. Put an empty
desiccant cartridge into the adapter on the waste bottle cap.
Note: The molecular sieve in the filled cartridges must be regenerated regularly depending on the humidity of the
laboratory. Refer to
Regenerate the molecular sieve
on page 19.
Push the inlet tube through the connector. Make sure that the end of the tube is at the bottom of the
bottle. Tighten the connector on the bottle cap.
Pump 1 is used to fill the measurement cell with solvent. Pump 2 is used to empty the measurement
cell. Refer to the illustrated steps that follow.
Tidy the work area
Attach the tubes to the instrument with the clips on the electrovalve and the sensor holder. Refer to
the illustrated steps that follow.
English
13
Содержание TitraLab KF1000 Series
Страница 2: ...English 3 Fran ais 23 Espa ol 45 Portugu s 67 88 107 128 2...
Страница 21: ...English 21...
Страница 42: ...42 Fran ais...
Страница 86: ...86 Portugu s...
Страница 90: ...4 5 1 1 1 6 11 2 7 12 2 3 8 13 1 4 USB 9 14 5 10 1 KF1121 1 2 90...
Страница 91: ...2 USB USB USB USB 2 1 24V 4 7 USB 2 1 5 8 3 2 6 3 1 3 3x 5 2 4 91...
Страница 92: ...4 1 KF 6 2 KF Teflon 7 3 10x 8 4 USB 9 3 x GL45 5 10 3x 1 1 2 1 15 6 GFCI GFI 92...
Страница 93: ...1 2 2 91 3 180 93...
Страница 94: ...1 90 94...
Страница 95: ...95...
Страница 96: ...96...
Страница 97: ...MSDS SDS 97...
Страница 98: ...104 1 2 98...
Страница 99: ...5 5 1 4 2 5 3 2 99...
Страница 100: ...MSDS SDS 1 Settings 2 Select 100 100 3 Back 1 Settings 2 3 4 5 100...
Страница 103: ...1 Data log 2 Select USB 3 4 Preview 1 calibration 2 3 Start 4 5 6 Continue 7 Yes No 8 Exit 101 103...
Страница 104: ...300 572 F 4 Maintenance 1 Syringe activation Syringe replacement 93 1 Pump activation Start 2 Toggle 104...
Страница 105: ...105...
Страница 106: ...1 Select Reset Edit 106...
Страница 109: ...1 2 3 4 5 1 1 1 6 11 2 7 12 2 3 8 13 1 4 USB 9 14 5 10 1 KF1121 1 2 109...
Страница 110: ...2 USB USB USB USB 2 1 24 V 4 7 USB 2 1 5 8 3 2 6 HACH Japan 3 1 3 3 5 2 4 110...
Страница 111: ...4 1 KF 6 2 KF Teflon 7 3 10 8 4 USB 9 GL45 3 5 10 3 1 1 2 1 15 cm 6 111...
Страница 112: ...AC GFCI GFI 1 2 2 110 3 2 1 2 180 2 112...
Страница 113: ...1 109 113...
Страница 114: ...114...
Страница 115: ...115...
Страница 116: ...MSDS SDS 116...
Страница 117: ...124 1 2 117...
Страница 118: ...5 5 1 4 2 5 3 2 118...
Страница 119: ...MSDS SDS 1 Settings 2 Select 119 5 PC 119 3 Back 1 Settings 2 Applications Edit 3 Edit 119...
Страница 122: ...19 Next 122 20 Exit 1 Cell management 2 1 1 2 2 3 Exit 1 Data log 2 Select USB 3 122...
Страница 123: ...4 Preview 1 Calibration 2 3 Start 4 5 6 Continue 7 Yes No 8 Exit 120 123...
Страница 124: ...300 572 F 4 Maintenance 1 Syringe activation Syringe replacement 112 1 Pump activation 2 Toggle 2 124...
Страница 125: ...125...
Страница 127: ...Excess of titrant Excess of water 127...
Страница 130: ...1 2 3 4 5 1 1 1 6 11 2 7 12 2 3 8 13 1 4 USB 9 14 5 10 1 KF1121 1 2 130...
Страница 131: ...2 USB USB USB USB 2 1 24V 4 7 USB 2 1 5 8 3 2 6 3 1 3 3x 5 2 4 131...
Страница 132: ...4 1 KF 6 2 KF Teflon 7 3 10x 8 4 USB 9 3 x GL45 5 10 3x 1 1 2 1 15cm 6 132...
Страница 133: ...AC GFCI GFI 1 2 2 131 3 180 133...
Страница 134: ...1 130 134...
Страница 135: ...135...
Страница 136: ...136...
Страница 137: ...MSDS SDS 137...
Страница 138: ...144 1 2 138...
Страница 139: ...5 5 1 4 2 5 3 2 139...
Страница 140: ...MSDS SDS 1 2 140 PC 140 3 1 2 3 4 5 140...
Страница 141: ...USB 147 1 2 USB USB USB 3 4 5 USB 1 2 3 4 2 5 6 MSDS SDS 141...
Страница 142: ...1 2 3 USB 4 142 5 6 7 8 9 10 11 12 13 14 11 15 16 17 18 14 19 142 20 142...
Страница 143: ...1 2 1 1 2 2 3 1 2 USB 3 4 1 2 3 4 5 6 7 8 143...
Страница 144: ...Purge 141 300 C 572 F 4 1 144...
Страница 145: ...133 1 2 145...
Страница 146: ...1 146...
Страница 147: ...147...
Страница 148: ......
Страница 149: ......