7.  

With the tip in contact with the side of the receiving vessel, slowly and 

smoothly press down on the cap until the turret reaches the stop and the solution 
is completely discharged. 

 

 

 

Maintenance 

Periodically disassemble the lower body.  Clean and lubricate the piston and 
retainer to ensure smooth operation.  

See Figure 1.

  During reassembly, apply a 

light coat of stopcock grease to the piston shaft and o-ring seal.  Slide the 
coupling over the housing extension and proceed with reassembly.  Tighten the 
coupling hand-tight only. 
 
If the TenSette Pipet is to be used at an altitude significantly different than 1000 
feet (305 meters) above sea level, the delivery volume will be affected. 

 See 

Figure 2.  

The instrument can be reset for greater accuracy as follows: 

 
1.  Remove the turret cap screw, turret cap and volume-setting ring.  Loosen the 

turret set screw by one or two turns. 

 
2.  Place a narrow-necked vessel (for example, a 25-mL volumetric flask) on an 

analytical balance and zero the balance (or record the weight if the balance 
is non zeroing). 

 
3.  Weigh the quantity of water dispensed at the 0.10-mL setting.  Use only 

high-quality demineralized water at 20-25ºC. 

 
4.  If the weight of the water is other than 0.10 g, use a small screwdriver to 

give the turret adjustment screw (inside the metering turret) a slight turn.  
Turn clockwise to increase delivery volume. 

 
5.  Zero the balance (or record the weight) just prior to adding another 0.10 mL 

of demineralized water.  If the added weight is other than 0.10 g, repeat 
Steps 3 through 5. 

Содержание TenSette 19700-01

Страница 1: ...19700 88 TenSette Pipet Model 19700 01 Hach Company 1983 1998 All rights are reserved 5 12 93 9ED Rev 1 10 98...

Страница 2: ......

Страница 3: ...ts of solutions For best results always use a new tip for each pipetting operation After being used several times the pipet tip may retain some liquid causing an error in delivery Each pipet is suppli...

Страница 4: ...ned 2 Turn the turret cap to align the desired volume on the volume setting ring with the mark on the housing assembly 3 Press down on the turret cap with the thumb using a smooth motion until the tur...

Страница 5: ...th the turret up withdraw the tip from the liquid and move it to the receiving vessel Avoid placing pressure on the cap while moving the pipet 6 Use the thumb and forefinger to twist the turret cap to...

Страница 6: ...4 Figure 1 TenSette Pipet Exploded View...

Страница 7: ...5...

Страница 8: ...volume will be affected See Figure 2 The instrument can be reset for greater accuracy as follows 1 Remove the turret cap screw turret cap and volume setting ring Loosen the turret set screw by one or...

Страница 9: ...just the delivery volume as described in Step 4 When the pipet delivers 1 00 g at the 1 00 mL setting check the 0 10 mL setting again 8 When the delivery volume is accurate for both the 1 00 mL and th...

Страница 10: ...les Service Canada Ltd 100 Dayton Ave 1313 Border Street Unit 34 P O Box 907 Winnipeg Manitoba Ames Iowa 50010 R3H 0X4 800 227 4224 U S A only 800 665 7635 FAX 515 232 1276 204 632 5598 Telephone 515...

Страница 11: ...tion Manual each 19707 00 Metering Turret each 19716 00 Pin 0 062 0 75 each 21856 96 Pipet Tips pkg of 50 each 19709 00 Piston Assembly each 19907 00 Retainer each 19717 00 Retaining Ring each 5651 00...

Страница 12: ...389 Telephone 970 669 3050 FAX 970 669 2932 FOR TECHNICAL ASSISTANCE PRICE INFORMATION AND ORDERING Call 800 227 4224 Contact the HACH office or distributor serving you www hach com techhelp hach com...

Отзывы: