Specification
Details
Operating temperature
5 to 40 °C (32 to 104 °F)
Storage temperature
-40 to 60 °C (-40 to 140 °F)
Operating humidity
Relative humidity: 5–95% at different temperatures, non-condensing
Sampling rate
New sample: every 1.9 minutes ±5% at 60 Hz; 2.3 minutes ±5% at
50 Hz
Sample flow rate to sample conditioning 50 to 500 mL/minute flow rate necessary (250 mL/minute
recommended)
Inlet pressure to instrument
1 to 5 psig (0.07 to 0.34 bar), 1.5 psig (0.1 bar) is optimum, > 5 psig
(0.34 bar) can cause sample tubing failure
Inlet pressure to sample conditioning
1.5 psig to 75 psig (0.1 to 5.2 bar)
Inlet fitting
At instrument, 6.35 mm (0.25 in.) OD polyethylene tubing with quick-
disconnect fitting
Drain fitting
Hose barb for 12.7 mm (0.5 in.) ID flexible tubing
Sample temperature range
5 to 40 °C (41 to 104 °F)
Certifications
CE, cETLus
Warranty
1 year (EU: 2 years)
Trip point specifications
Alarm trip point
Minimum trip value
Maximum trip value
Temperature influence on trip point
0.3 mg/L
0.22 mg/L
0.38 mg/L
–0.03 mg/L per °C
1 mg/L
0.75 mg/L
1.25 mg/L
–0.03 mg/L per °C
2 mg/L
1.5 mg/L
2.5 mg/L
–0.03 mg/L per °C
5 mg/L
3.75 mg/L
6.25 mg/L
–0.06 mg/L per °C
10 mg/L
7.5 mg/L
12.5 mg/L
–0.08 mg/L per °C
20 mg/L
15.0 mg/L
25.0 mg/L
–0.09 mg/L per °C
50 mg/L
37.5 mg/L
62.5 mg/L
–0.29 mg/L per °C
100 mg/L
75.0 mg/L
125.0 mg/L
–0.60 mg/L per °C
General information
In no event will the manufacturer be liable for direct, indirect, special, incidental or consequential
damages resulting from any defect or omission in this manual. The manufacturer reserves the right to
make changes in this manual and the products it describes at any time, without notice or obligation.
Revised editions are found on the manufacturer’s website.
Safety information
N O T I C E
The manufacturer is not responsible for any damages due to misapplication or misuse of this product including,
without limitation, direct, incidental and consequential damages, and disclaims such damages to the full extent
permitted under applicable law. The user is solely responsible to identify critical application risks and install
appropriate mechanisms to protect processes during a possible equipment malfunction.
Please read this entire manual before unpacking, setting up or operating this equipment. Pay
attention to all danger and caution statements. Failure to do so could result in serious injury to the
operator or damage to the equipment.