Hach SP-510 Скачать руководство пользователя страница 24

Cause

Solution

If the flow is too low, the
sample cell does not flush
completely all the color out of
the colorimeter. This causes a
zero reading. If the flow is too
high, some of the water will
bypass the pinch block and
cause the color to be diluted.

Make sure to set the flow rate to 200 mL/minute.

The sample line is pinched off
in the pinch block.

• Pull the sample line from the colorimeter. Water that is already in the line

could leak out. Examine and adjust the flow rate when a steady water
stream comes out while the line is pinched. Set the flow rate to
200 mL/minute.

• Make sure that the pressure plate of the pinch block is not overtightened.
• Examine the back of the pinch blocks and make sure that they do not have

grooves.

No stir bar is installed or more
than one stir bar is installed.

• Make sure that a stir bar is installed.
• Use a paper clip to retrieve the stir bar from the colorimeter and make sure

that multiple stir bars are not installed.

• If the stir bar does not operate, mix the sample manually after the reagents

have been added. Use the end of a wooden Q-tip and mix the reagents
approximately for 10 seconds. If the instrument gets a reading, the mixing is
not sufficient. Replace the stir coil. Refer to 

Replacement parts and

consumables

 on page 22.

The instrument does not
receive any buffer or indicator
solution.

Remove the two reagent lines from the Y connector. Only one drop of reagent
comes out per cycle. If no reagent comes out, examine the pressure plate.
Make sure that the pinch block is not overtightened and the reagent lines are
not pinched. Examine the back of the pinch block for grooves. Replace the
pinch block if necessary.

The chemistry does not operate
correctly.

Mix 1-mL of each reagent and 80-mL of sample. A color change must show. If
no color change shows, replace the reagents.

The instrument does not
operate correctly.

Pinch the sample inlet line to stop the sample flow. The next cycle reads soft. If
it does not read soft, examine the LED if it shows an orange light. Examine the
photocell and clean if necessary. After this cycle, the instrument reads hard
again.

Replacement parts and consumables

W A R N I N G

 

Personal injury hazard. Use of non-approved parts may cause personal injury, damage to the
instrument or equipment malfunction. The replacement parts in this section are approved by the
manufacturer.

Note: Product and Article numbers may vary for some selling regions. Contact the appropriate distributor or refer to
the company website for contact information.

Refer to the installation kit and maintenance kit documentation for more information.

Replacement parts

Description

Item no.

Cam/coupler assembly

5410500 

Circuit board assembly

5694800 

Colorimeter flow-cell assembly

5516200 

22

   

English

Содержание SP-510

Страница 1: ...DOC023 53 80422 SP 510 01 2017 Edition 8 User Manual...

Страница 2: ......

Страница 3: ...12 Electrical connections for conduit 13 Wiring for power 14 Select the voltage 14 Connect the alarm relays 14 Install the buffer solution 15 Install the indicator solution 15 Install the stir bar 16...

Страница 4: ...Replace the fuse 20 Troubleshooting 21 Troubleshooting for a hard reading 21 Replacement parts and consumables 22 Index 25 Table of Contents 2...

Страница 5: ...ts Buffer and indicator solution 500 mL of each reagent every 2 months Reagent shelf life 1 year Power requirements 115 230 VAC 70 VA 50 60 Hz 1 25 A Fuse Electrical connections Two 3 wire barrier ter...

Страница 6: ...nformation N O T I C E The manufacturer is not responsible for any damages due to misapplication or misuse of this product including without limitation direct incidental and consequential damages and...

Страница 7: ...ock and or electrocution exists This symbol indicates that a risk of fire is present This symbol when noted on the product identifies the location of a fuse or current limiting device Certification Ca...

Страница 8: ...he above Product overview D A N G E R Chemical or biological hazards If this instrument is used to monitor a treatment process and or chemical feed system for which there are regulatory limits and mon...

Страница 9: ...9 Indicator and reagent bottles 4 Air purge optional 10 Sample inlet 5 Access cover 11 Enclosure drain 6 Colorimeter 12 Sample drain Product components Make sure that all components have been received...

Страница 10: ...lified personnel must conduct the tasks described in this section of the document Install the instrument Install this instrument on an indoor wall away from direct sunlight For a complete purge of the...

Страница 11: ...Figure 3 Dimensions for wall installation Figure 4 Door clearance Plumb the instrument C A U T I O N Fire hazard This product is not designed for use with flammable liquids English 9...

Страница 12: ...air purge optional An air purge is necessary in an environment with high humidity or caustic vapors Note Use only dry oil free instrument air at 0 42 m3 15 standard cubic feet per hour scfh 1 Find th...

Страница 13: ...tall the pressure plate refer to Figure 6 Figure 6 Install the pressure plate Install the sample conditioning kit The sample conditioning kit is supplied with the instrument The kit removes large part...

Страница 14: ...ion point 8 Low flow valve option 3 Drainage flow 6 Filtered sample bypass ball valve shown open 9 High flow valve option Electrical installation D A N G E R Electrocution hazard Always remove power t...

Страница 15: ...power source W A R N I N G Electrocution hazard Install a 10 A circuit breaker for mains power Identify the circuit breaker with a label as a local disconnect for this equipment The instrument has pl...

Страница 16: ...slide the voltage switch to the 230 V position Refer to Figure 10 on page 14 Note The fuses are approved for North American and European use and do not need to be changed with the voltage Connect the...

Страница 17: ...ory and are ready to install without preparation Put the bottle in the instrument as shown in Figure 12 on page 15 Refer to Buffer and indicator solution for the applicable trip point on page 24 Insta...

Страница 18: ...e 3 for the keypad description and navigation information Figure 14 SP 510 keypad Table 3 Keypad description Key Function Description 1 HARD LED The monitor found hardness in the water The status indi...

Страница 19: ...gents to decrease the start up time to 15 minutes 1 Set the sample flow to the instrument to off 2 Disconnect the reagent feed line after the Y fitting at the colorimeter 3 Connect a syringe to the tu...

Страница 20: ...a stable green light The HARD LED comes on with a red light when the sample reagent mixture measurement is higher than the trip point for one measurement Push ALARM OFF to set the alarm relay to off T...

Страница 21: ...te Do not pour the remaining reagent into new bottles One 500 mL bottle of each buffer and indicator solution can be used for approximately 2 months 1 Discard the old containers with remaining content...

Страница 22: ...uric acid standard solution in the colorimeter Note Lower normality sulfuric acid is not sufficient to clean the colorimeter 4 Let the sulfuric acid stay in the colorimeter for 15 minutes 5 Use a wood...

Страница 23: ...not connected Connect the motor cable connection Motor does not operate Replace the motor The reading is continually high The HARD LED is on No stir bar LED is not connected No sample flow Out of reag...

Страница 24: ...instrument does not receive any buffer or indicator solution Remove the two reagent lines from the Y connector Only one drop of reagent comes out per cycle If no reagent comes out examine the pressur...

Страница 25: ...in ID white 4271700 Tubing 0 11 cm 0 043 in ID brown 5412100 Y fitting strainer 40 mesh 4661600 Accessories Description Item no Seal oil tight 4221000 Power cord kit for 115 V North American operation...

Страница 26: ...rip point Buffer5 item no Indicator item no 0 3 mg L 2768549 2794649 1 mg L 2768549 2769049 2 mg L 2768549 2769149 5 mg L 2768549 2769249 10 mg L 2768649 2769249 20 mg L 2768749 2769249 50 mg L 276884...

Страница 27: ...1 Colorimeter 20 D Dimensions 8 F Fuse 20 H Hard calibration 17 Hard reading 21 I Indicator solution 15 K Keypad 16 N Navigation 16 P Pressure plate 11 R Reagent 15 19 Reagent line 21 Relay 14 S Sampl...

Страница 28: ......

Страница 29: ......

Страница 30: ...s hach com www hach com HACH LANGE GMBH Willst tterstra e 11 D 40549 D sseldorf Germany Tel 49 0 2 11 52 88 320 Fax 49 0 2 11 52 88 210 info de hach com www de hach com HACH LANGE S rl 6 route de Comp...

Отзывы: