Cause
Solution
If the flow is too low, the
sample cell does not flush
completely all the color out of
the colorimeter. This causes a
zero reading. If the flow is too
high, some of the water will
bypass the pinch block and
cause the color to be diluted.
Make sure to set the flow rate to 200 mL/minute.
The sample line is pinched off
in the pinch block.
• Pull the sample line from the colorimeter. Water that is already in the line
could leak out. Examine and adjust the flow rate when a steady water
stream comes out while the line is pinched. Set the flow rate to
200 mL/minute.
• Make sure that the pressure plate of the pinch block is not overtightened.
• Examine the back of the pinch blocks and make sure that they do not have
grooves.
No stir bar is installed or more
than one stir bar is installed.
• Make sure that a stir bar is installed.
• Use a paper clip to retrieve the stir bar from the colorimeter and make sure
that multiple stir bars are not installed.
• If the stir bar does not operate, mix the sample manually after the reagents
have been added. Use the end of a wooden Q-tip and mix the reagents
approximately for 10 seconds. If the instrument gets a reading, the mixing is
not sufficient. Replace the stir coil. Refer to
Replacement parts and
consumables
on page 22.
The instrument does not
receive any buffer or indicator
solution.
Remove the two reagent lines from the Y connector. Only one drop of reagent
comes out per cycle. If no reagent comes out, examine the pressure plate.
Make sure that the pinch block is not overtightened and the reagent lines are
not pinched. Examine the back of the pinch block for grooves. Replace the
pinch block if necessary.
The chemistry does not operate
correctly.
Mix 1-mL of each reagent and 80-mL of sample. A color change must show. If
no color change shows, replace the reagents.
The instrument does not
operate correctly.
Pinch the sample inlet line to stop the sample flow. The next cycle reads soft. If
it does not read soft, examine the LED if it shows an orange light. Examine the
photocell and clean if necessary. After this cycle, the instrument reads hard
again.
Replacement parts and consumables
W A R N I N G
Personal injury hazard. Use of non-approved parts may cause personal injury, damage to the
instrument or equipment malfunction. The replacement parts in this section are approved by the
manufacturer.
Note: Product and Article numbers may vary for some selling regions. Contact the appropriate distributor or refer to
the company website for contact information.
Refer to the installation kit and maintenance kit documentation for more information.
Replacement parts
Description
Item no.
Cam/coupler assembly
5410500
Circuit board assembly
5694800
Colorimeter flow-cell assembly
5516200
22