71
Section 8
Calibration
8.1 Calibration
The transmitter is adjusted to the sensor through the calibration. The
available calibration methods are shown in
. Calibrate the
sensor in air when the sensor can be removed from the process. The
sensor will stabilize more quickly in air.
When the sensor cannot be removed from the process (e.g. sterile
processes), the process medium must be aerated during calibration
(e.g. after sterilization).
During calibration the transmitter remains in the Hold mode for safety
reasons. The loop current is frozen at the value specified in the
o1.HoLD menu option. The Sensoface icon is inactive. The calibration
mode indicator is displayed (
).
To exit the calibration mode at any time, press
CAL
and then
ENTER
.
The output current will be held for 20 seconds and the measured value
will be displayed.
Note: Calibrations must be performed by trained personnel. Incorrectly
set parameters may result in errors in sample measurements.
Table 21 Calibration methods and passcodes
Method
Passcode
Zero calibration
1001
Temperature probe adjustment
1015
Zero/slope display
0000
Calibration in air or water
1100
Calibration by product comparison
1105
Содержание si792 D series
Страница 2: ...2 ...
Страница 6: ...Table of Contents 4 ...
Страница 12: ...Specifications 10 ...
Страница 36: ...Interface and navigation 34 ...
Страница 44: ...Operation si792 x D 4 20 mA HART 42 ...
Страница 66: ...Operation Foundation Fieldbus 64 ...
Страница 72: ...Operation Profibus PA 70 ...
Страница 80: ...Maintenance 78 ...
Страница 86: ...Troubleshooting 84 ...
Страница 88: ...Parts and accessories 86 ...
Страница 93: ...Index 91 Z zero adjustment 72 ...
Страница 94: ...Index 92 ...
Страница 95: ...93 ...