SECTION 9
REFERENCE TABLES
The following tables have been provided as a reference, but are
not required for use with the DO meter.
9.1
Salinity Correction Factors
Use the values in Table 7 if the conductivity meter in use does not
measure salinity. Use a conductivity meter to obtain conductivity
in mS/cm at reference temperature (20 °C), then use Table 7 to
estimate the salinity correction factor (in ppt*) to the nearest
whole number. Enter the salinity value from Table 7 into the
meter per setup function Section 3.1.4 on page 24.
This table was calculated up to the conductivity of 54 mmhos/cm
from the International Oceanographic Tables**.
*Salinity determined by the conductivity at 20 °C.
* ppt = Parts per Thousands of Salinity
** International Oceanographic Tables, Vol. I, National Institute of Oceanography of Great Britain,
Womley, Godaming, Surrey, England and Uncesco, Paris 1971.
Table 7 Salinity Correction Factors
Conductivity in
mS/cm
Salinity
value*
Conductivity in
mS/cm
Salinity
value*
Conductivity in
mS/cm
Salinity
value*
5
3
20
13
35
25
6
4
21
14
36
25
7
4
22
15
37
26
8
5
23
15
38
27
9
6
24
16
39
28
10
6
25
17
40
29
11
7
26
18
42
30
12
8
27
18
44
32
13
8
28
19
46
33
14
9
29
20
48
35
15
10
30
21
50
37
16
10
31
22
52
38
17
11
32
22
54
40
18
12
33
23
—
—
19
13
34
24
—
—
Содержание sension8
Страница 6: ...6 ...
Страница 8: ...8 ...
Страница 10: ...10 ...
Страница 12: ...12 ...
Страница 20: ...20 ...
Страница 44: ...44 ...
Страница 48: ...48 ...
Страница 50: ...50 ...
Страница 64: ...64 ...
Страница 76: ...76 ...
Страница 78: ...78 HOW TO ORDER ...
Страница 79: ...79 REPAIR SERVICE ...
Страница 84: ......