8.
Push to measure the second calibration solution.
The next calibration solution is shown.
9.
Rinse the probe with deionized water and put the probe into the third
calibration beaker. Be sure that there are no air bubbles in the
membrane.
10.
Push to measure the third calibration solution.
When the calibration is good, the display briefly shows Calibration
OK and then returns to the main menu.
Note: When a printer is connected the print menu opens and the result can be
printed.
View the calibration data
Data from the most recent calibration can be shown.
1.
From the main menu use the or to select DATA LOGGER.
Confirm.
2.
Select Display data.
3.
Select Calibration data and confirm with . The last calibration data
is shown.
• pH-the slope and offset values are shown alternating with the
deviation (in %) and calibration temperature.
• ORP-the measured mV value and calibration temperature are
shown.
• Conductivity-the cell constant and calibration temperature for each
standard are shown.
Set the calibration reminder
pH:
The calibration reminder can be set between 0 to 23 hours or
1-7 days (default 1 day).
EC:
The calibration reminder can be set
between 0 to 99 days (default 15 days). The display shows the
remaining time to the new calibration.
Note: When 0 days is selected, the calibration reminder is turned off.
1.
From the main menu use the or to select CALIBRATION.
Confirm.
2.
Use the to enter the calibration menu.
3.
Use the or to select Cal. frequency and confirm.
4.
Use the and to advance to the next step and use the or to
change a value. Confirm.
Push to start the calibration.
Sample measurements
Each probe has specific preparation steps and procedures for taking
sample measurements.
1.
From the main menu use the or and and to select
MEASURE. Confirm.
2.
Use the to change the following settings. Confirm every entry.
Option
Description
Resolution
Select the resolution: 1, 0.1, 0.01 (default) or 0.001
Measure
Stability—select By stability Criterion: Fast (variation <
0.02 pH in 6 s), Standard (variation < 0.01 pH in 6 s) or Strict
(variation < 0.002 pH in 6 s). In continuous—enter the time
interval for the In continuous Acquis. interval (data storage or
printing data). By time—enter the time interval for data storage
or printing data.
Display mV
Display mV—select YES or NO to show mV.
Limits
Limits—select YES or NO. YES: Enter the upper and lower
limits. An acoustic warning appears when the measurement is
out of limit. The report output shows an A next to the
measured value when the measurement was out of limit.
Isopotential
Isopotential—change the Isopotential pH value in Data
introduction. Select Calculate to calibrate the probe again.
TC
TC—select Linear or Natural waters. Linear: enter a value in
%/°C (default 2.00%/Temperature). Natural waters: Non-linear
for natural waters, according to EN27888
Tref
Reference temperature—select between 20 or 25 °C or Other
temperature.
3.
Push to start the measurement.
Note: If the measurement is not stabilizing after 120 seconds, the meter turns
automatically into the continuous measurement mode.
English
11
Содержание sensION+ MM374
Страница 2: ...English 3 Français 23 Español 44 Português 65 中文 85 日本語 103 한글 123 ไทย 143 2 ...
Страница 6: ...1 2 3 4 6 English ...
Страница 16: ...1 2 16 English ...
Страница 26: ...1 2 3 4 26 Français ...
Страница 37: ...1 2 Français 37 ...
Страница 47: ...1 2 3 4 Español 47 ...
Страница 58: ...1 2 58 Español ...
Страница 68: ...1 2 3 4 68 Português ...
Страница 78: ...1 2 78 Português ...
Страница 88: ...1 2 3 4 88 中文 ...
Страница 97: ...1 2 中文 97 ...
Страница 106: ...1 2 3 4 106 日本語 ...
Страница 116: ...1 2 116 日本語 ...
Страница 126: ...1 2 3 4 126 한글 ...
Страница 136: ...1 2 136 한글 ...
Страница 146: ...1 2 3 4 146 ไทย ...
Страница 156: ...1 2 156 ไทย ...
Страница 163: ......