3.
Use the or to select
Option
Description
Automatic
A consecutive number will be automatically assigned to every
sample.
Manual
A keyboard or a barcode scanner is required to enter the sample
ID name before taking a measurement (maximum 15 characters).
Select the data output
Data can be stored or transferred to a printer or to a PC.
1.
From the main menu use the or to select SYSTEM. Confirm.
2.
Use the or to select Data Output and confirm.
3.
Use the or to select
Option
Description
Deactivated
Select Deactivated if no printer or PC is connected.
For Printer
Select Dot matrix printer or Thermal printer.
For Computer
Select Terminal, ComLabo or ComLabo Easy. The ComLabo
Software controls several modules, pH and conductivity
meters, automatic burettes, Samplers and so on from a
computer. The ComLabo Easy software gets pH and
conductivity data from a PC.
Change the date and time
The date and time can be changed from the Date / Time menu.
1.
From the main menu use the or to select SYSTEM. Confirm.
2.
Use the or to select Date / Time and confirm.
3.
Use the and to advance to the next step and use the or to
change a value. Confirm.
The current date and time will be shown on the display.
Adjust the display contrast
1.
From the main menu use the or to select SYSTEM. Confirm.
2.
Use the or to select Display contrast and confirm.
3.
Use the and to adjust the contrast of the display and confirm.
Adjust the temperature
The temperature measurement can be adjusted at 25 °C (77 °F) and/or
85 °C (185 °F) to increase accuracy.
1.
Put the probe and a reference thermometer in a container of water at
approximately 25 °C and allow the temperature to stabilize.
2.
Compare the temperature read by the meter with that of the reference
thermometer. The difference is the adjustment value for the meter.
Example: reference thermometer: 24.5 °C; meter: 24.3 °C. Adjustment
value: 0.2 °C.
3.
Enter the adjustment value for the 25 °C reading:
a.
From the main menu use the or to select SYSTEM. Confirm.
b.
Use the or to select Readjust temp. and confirm.
c.
Use the or to select 25 °C and confirm.
d.
Use the arrow keys to enter the adjustment value for 25 °C.
Confirm.
4.
Put the probe and a reference thermometer in a container of water at
approximately 85 °C and allow the temperature to stabilize.
5.
Compare the temperature from the meter with that of the reference
thermometer. The difference is the adjustment value for the meter.
a.
Use the or to select 85 °C and confirm.
b.
Use the arrow keys to enter the adjustment value for 85 °C.
Confirm.
c.
Select Save changes and confirm.
Change the stirring settings
The magnetic stirrer can be turned on and the stirring speed can be
changed in the Stirring menu.
10
English
Содержание sensION+ EC71
Страница 2: ...English 3 Français 17 Español 32 Português 47 中文 62 日本語 76 한글 91 ไทย 105 2 ...
Страница 5: ...1 2 3 4 English 5 ...
Страница 13: ...1 2 English 13 ...
Страница 19: ...1 2 3 4 Français 19 ...
Страница 28: ...1 2 28 Français ...
Страница 34: ...1 2 3 4 34 Español ...
Страница 43: ...1 2 Español 43 ...
Страница 49: ...1 2 3 4 Português 49 ...
Страница 58: ...1 2 58 Português ...
Страница 64: ...1 2 3 4 64 中文 ...
Страница 72: ...1 2 72 中文 ...
Страница 78: ...1 2 3 4 78 日本語 ...
Страница 87: ...1 2 日本語 87 ...
Страница 93: ...1 2 3 4 한글 93 ...
Страница 101: ...1 2 한글 101 ...
Страница 107: ...1 2 3 4 ไทย 107 ...
Страница 115: ...1 2 ไทย 115 ...
Страница 119: ......