![Hach sensION+ EC5 Скачать руководство пользователя страница 10](http://html2.mh-extra.com/html/hach/sension-ec5/sension-ec5_user-manual_3385223010.webp)
Startup
Turn the meter on and off
N O T I C E
Make sure that the probe is connected to the meter before the meter is turned on.
Push and hold to turn on or turn off the meter. If the meter does not
turn on, make sure that the batteries are properly installed.
Note: The meter is set to automatically power off after 5 minutes of inactivity to
maximize the battery life.
Operation
W A R N I N G
Chemical exposure hazard. Obey laboratory safety procedures and wear all of
the personal protective equipment appropriate to the chemicals that are handled.
Refer to the current material safety data sheets (MSDS) for safety protocols.
Setup for pH
The meter is ready for pH or ORP calibration without additional setup.
Setup for dissolved oxygen
Before calibration or measurement for dissolved oxygen, the probe must
be polarized and the settings for atmospheric pressure and salinity must
be entered.
Polarize the electrode
If the probe or batteries are disconnected, connect the probe or install
the batteries and wait for polarization:
Disconnect time
Polarization time
< 5 minutes
10 minutes
5 to 15 minutes
45 minutes
> 15 minutes
6 hours
Change the settings
The settings for dissolved oxygen measurements can be changed when
a dissolved oxygen probe is used. Push
to view the current settings.
1.
Push
. The first setting is shown.
2.
Quickly use the arrow keys to change the value. Wait for the next
setting to be shown and change any additional values:
Option
Description
Atmospheric pressure in mbar (default: 1013 mbar)
SAL
Salinity in g/L NaCl (default: 0 g/L)
Aut
: automatic (only in model MM156)
SEn
Select the DO sensor model:
5120
or
5130
Note: The salinity value can be found from a conductivity measurement. Refer
to
on page 17.
Setup for conductivity
The settings for conductivity measurements can be changed when a
conductivity probe is used. Push
to view the current settings.
1.
Push
. The first setting is shown.
2.
Quickly use the arrow keys to change the value. Wait for the next
setting to be shown and change any additional values:
Option Description
CEL
Electrode type: platinum (default) or titanium.
Note: This option is not available on all meters.
CAL
The number of standards for conductivity calibration. Select one
standard (default). Select the standard nearest to the expected
measurement range.
tC
Temperature compensation: 0 to 9.99%/°C (default: 2%/°C)
tr
Reference temperature: 20 or 25 °C (default: 25 °C)
F
TDS calculation factor: 0.01 to 4.44 (default: 0.64)
10
English
Содержание sensION+ EC5
Страница 194: ...TDS 0 00 o 4 44 0 64 5 CE 194...
Страница 196: ...1 2 3 4 5 NiMH NiMH 1 2 V 1 5 V 2300 mAH NiMH 90 NiMH AA 5 Display Options 196...
Страница 197: ...1 IP67 1 1 2 2 2 197...
Страница 198: ...3 1 4 2 5 3 3 4 pH1 1 pH ORP 4 2 5 3 198...
Страница 199: ...5 DO6 1 DO 4 2 5 3 6 MM110 1 pH 4 2 5 3 6 ORP 199...
Страница 200: ...7 MM150 1 5 2 pH 6 3 7 TDS ORP 4 8 8 MM156 1 5 2 pH 6 3 7 DO 4 8 200...
Страница 201: ...9 EC5 1 4 2 5 3 6 TDS 5 MSDS pH pH ORP 201...
Страница 203: ...1 2 3 4 5 6 pH 7 pH 8 9 10 11 1 2 pH ORP 1 2 3 pH ORP mV pH ORP 1 2 3 OFF 203...
Страница 205: ...5 1 2 OFF 3 OFF 1 2 2 3 3 2 205...
Страница 206: ...10 10 E1 206...
Страница 207: ...E2 250 nA E2 pH 70 105 E2 30 E3 0 05 1 207...
Страница 208: ...E3 58 mV E4 E4 TC 0 E5 E6 ErA MM156 50 g L 208...
Страница 259: ...0 00 4 44 0 64 5 CE 259...
Страница 261: ...2 3 4 5 NiMH NiMH 1 2 1 5 NiMH 2300 90 NiMH 5 261...
Страница 262: ...1 IP67 1 1 2 2 2 3 1 4 2 5 3 262...
Страница 263: ...3 4 pH1 1 pH 4 2 5 3 5 DO6 1 4 2 5 3 263...
Страница 264: ...6 MM110 1 pH 4 2 5 3 6 7 MM150 1 5 2 pH 6 3 7 4 8 264...
Страница 265: ...8 MM156 1 5 2 pH 6 3 7 4 8 9 EC5 1 4 2 5 3 6 265...
Страница 266: ...5 MSDS pH pH 5 10 5 15 45 15 6 1 2 1013 SAL NaCl 0 Aut MM156 SEn DO 5120 5130 3 274 266...
Страница 267: ...1 2 CEL CAL tC 0 9 99 C 2 C tr 20 25 C 25 C F 0 01 4 44 0 64 MSDS 1 2 3 4 5 6 pH 7 pH 8 9 10 11 OK 267...
Страница 269: ...5 6 85 C 2 1 2 C F 5 1 2 OFF no 3 OFF 1 2 2 3 3 2 269...
Страница 270: ...10 10 E1 270...
Страница 271: ...E2 250 E2 pH 70 105 E2 30 E3 0 05 1 271...
Страница 272: ...E3 58 E4 E4 TC 0 TC E5 E6 ErA MM156 50 TC 272...
Страница 275: ...3 NaCl 33 6 20 2 41 2 25 4 48 9 30 6 56 3 35 8 62 2 40 0 69 4 45 3 75 7 50 0 275...
Страница 338: ...TDS 0 00 4 44 0 64 5 CE 338...
Страница 340: ...3 4 5 NiMH NiMH 1 2 V 1 5 V 2300 mAH NiMH 90 NiMH 5 340...
Страница 341: ...1 IP67 1 1 2 2 2 3 1 4 2 5 ON OFF 3 341...
Страница 342: ...3 Measure Calibration Standby 4 pH1 1 pH ORP 4 2 5 3 5 DO6 1 DO 4 2 5 3 342...
Страница 343: ...6 MM110 1 pH 4 2 5 3 6 ORP 7 MM150 1 5 2 pH 6 3 7 TDS ORP 4 8 343...
Страница 344: ...8 MM156 1 5 2 pH 6 3 7 4 8 9 EC5 1 4 2 5 3 6 TDS 344...
Страница 345: ...5 MSDS pH pH ORP 5 10 5 15 45 15 6 1 2 mbar 1013 mbar SAL g L NaCl 0 g L Aut MM156 SEn DO 5120 5130 3 353 345...
Страница 346: ...1 2 CEL CAL tC 0 9 99 C 2 C tr 20 25 C 25 C F TDS 0 01 4 44 0 64 MSDS 1 2 3 4 5 6 pH 7 pH 8 9 10 11 OK 346...
Страница 348: ...5 6 85 C 2 1 2 C F 5 1 2 OFF no 3 OFF 1 2 2 3 3 2 348...
Страница 349: ...10 10 E 1 349...
Страница 350: ...E 2 250 mA E 2 pH 70 105 E2 30 E3 0 05 cm 1 350...
Страница 351: ...E3 58 mV E4 E4 TC 0 TC E5 E6 ErA MM156 50 g L TC 351...
Страница 354: ...3 mS cm g L NaCl 56 3 35 8 62 2 40 0 69 4 45 3 75 7 50 0 354...
Страница 386: ......
Страница 387: ......