Change the settings
The settings for dissolved oxygen measurements can be changed when
a dissolved oxygen probe is used. Push
to view the current settings.
1.
Push
. The first setting is shown.
2.
Quickly use the arrow keys to change the value. Wait for the next
setting to be shown and change any additional values:
Option
Description
Atmospheric pressure in mbar (default: 1013 mbar)
SAL
Salinity in g/L NaCl (default: 0 g/L)
Aut
: automatic (only in model MM156)
Note: The salinity value can be found from a conductivity measurement. Refer
to
on page 19.
Setup for conductivity
The settings for conductivity measurements can be changed when a
conductivity probe is used. Push
to view the current settings.
1.
Push
. The first setting is shown.
2.
Quickly use the arrow keys to change the value. Wait for the next
setting to be shown and change any additional values:
Option Description
CEL
Electrode type: platinum (default) or titanium.
Note: This option is not available on all meters.
CAL
The number of standards for conductivity calibration. Select one
standard (default). Select the standard nearest to the expected
measurement range.
tC
Temperature compensation: 0 to 9.99%/°C (default: 2%/°C)
tr
Reference temperature: 20 or 25 °C (default: 25 °C)
F
TDS calculation factor: 0.01 to 4.44 (default: 0.64)
The new settings are automatically stored in the meter.
Calibration
W A R N I N G
Chemical exposure hazard. Obey laboratory safety procedures and wear all of
the personal protective equipment appropriate to the chemicals that are handled.
Refer to the current material safety data sheets (MSDS) for safety protocols.
Calibration procedure
This procedure is for general use with liquid calibration solutions. Refer
to the documents that are included with each probe for additional
information.
1.
Pour the buffers or calibration solution into the labeled calibration
tubes.
2.
Rinse the probe with deionized water and put the probe into the
appropriate calibration tube. Make sure that there are no air bubbles
under the probe tip.
3.
Push . The parameter flashes.
4.
Use the arrow keys to change the parameter, if applicable.
5.
Push to select the parameter.
6.
Push the arrow keys to select the appropriate standard solution. For
pH, the buffers are automatically recognized.
7.
Push to measure the calibration solution.
Note: For pH, the next calibration solution is shown. For conductivity, use one
calibration point. Select the standard nearest to the expected measurement
range.
8.
Rinse the probe with deionized water and put the probe into the
second calibration tube, if applicable. Make sure that there are no air
bubbles under the probe tip.
9.
Push to measure the second calibration solution, if applicable.
The next calibration solution is shown.
10.
Rinse the probe with deionized water and put the probe into the third
calibration tube. Make sure that there are no air bubbles under the
probe tip.
11.
Push to measure the third calibration solution, if applicable.
English
11
Содержание sensION+ EC5 DL
Страница 2: ...English 3 Fran ais 20 Espa ol 40 Portugu s 58 77 93 111 128 2...
Страница 80: ...1 IP67 1 1 2 2 2 3 1 5 2 6 3 7 4 80...
Страница 81: ...3 4 PH1 DL 1 4 2 5 3 6 pH ORP 5 DO6 DL 1 4 2 5 3 6 DO 81...
Страница 82: ...6 MM110 DL 1 5 2 6 ORP 3 7 pH 4 7 MM150 DL 1 6 2 pH 7 TDS ORP 3 8 4 9 5 82...
Страница 83: ...8 MM156 DL 1 6 2 pH 7 DO 3 8 4 9 5 9 EC5 DL 1 5 2 6 TDS 3 7 4 83...
Страница 88: ...10 portCom USB RF 1 USB RF USB USB RF 2 portCom 3 Snd portCom 11 4 5 6 txt csv 11 88...
Страница 89: ...12 IP67 12 E1 E2 250 nA E2 pH 70 105 E2 30 89...
Страница 96: ...1 IP67 1 1 2 2 2 3 1 5 2 6 3 7 4 96...
Страница 97: ...3 4 PH1 DL 1 4 hh mm ss 2 5 3 6 pH ORP 5 DO6 DL 1 4 hh mm ss 2 5 3 6 DO 97...
Страница 98: ...6 MM110 DL 1 5 2 6 ORP 3 7 pH 4 hh mm ss 7 MM150 DL 1 6 2 pH 7 TDS ORP 3 8 4 9 5 hh mm ss 98...
Страница 99: ...8 MM156 DL 1 6 2 pH 7 DO 3 8 4 9 5 hh mm ss 9 EC5 DL 1 5 hh mm ss 2 6 TDS 3 7 4 1 99...
Страница 100: ...5 MSDS SDS EMI 20 cm 20 cm pH pH ORP 5 10 5 15 45 15 6 100...
Страница 105: ...3 2 Snd portCom 11 4 Save 5 6 txt csv Save 11 3 105...
Страница 106: ...12 IP67 12 E1 E2 250 nA E2 pH 70 105 106...
Страница 107: ...E2 30 E3 0 05 cm 1 E3 58 mV E4 E4 TC 0 TC E5 E6 ErA MM156 50 g L TC 107...
Страница 110: ...3 mS cm g L NaCl 41 2 25 4 48 9 30 6 56 3 35 8 62 2 40 0 69 4 45 3 75 7 50 0 110...
Страница 114: ...1 IP67 1 1 2 2 2 3 1 DISPLAY LIGHT 5 DOWN 2 DISPLAY LIGHT 6 ON OFF 3 UP 7 CALIBRATION 4 MEASUREMENT 114...
Страница 115: ...3 4 PH1 DL 1 4 hh mm ss 2 5 3 6 pH ORP 5 DO6 DL 1 4 hh mm ss 2 5 3 6 DO 115...
Страница 116: ...6 MM110 DL 1 5 2 6 ORP 3 7 pH 4 hh mm ss 7 MM150 DL 1 6 2 pH 7 TDS ORP 3 8 4 9 5 hh mm ss 116...
Страница 117: ...8 MM156 DL 1 6 2 pH 7 DO 3 8 4 9 5 hh mm ss 9 EC5 DL 1 5 hh mm ss 2 6 TDS 3 7 4 117...
Страница 118: ...5 MSDS SDS EMI 20 cm 7 9 20cm 7 9 in pH pH ORP 5 10 5 15 45 15 6 118...
Страница 123: ...11 123...
Страница 124: ...12 IP67 12 E1 E2 250nA E2 pH 70 105 124...
Страница 125: ...E2 30 E3 0 05cm 1 E3 58mV E4 E4 TC 0 TC E5 E6 ErA MM156 50 g L TC E7 E8 E9 USB RF portCom 125...
Страница 131: ...1 IP67 1 1 2 2 2 3 1 DISPLAY LIGHT 5 2 DATALOG 6 ON OFF 3 7 CALIBRATION 4 MEASUREMENT 131...
Страница 132: ...3 Measure Calibration Standby 4 PH1 DL 1 4 2 5 3 6 pH ORP 5 DO6 DL 1 4 2 5 3 6 DO 132...
Страница 133: ...6 MM110 DL 1 5 2 6 ORP 3 7 pH 4 7 MM150 DL 1 6 2 pH 7 TDS ORP 3 8 4 9 5 133...
Страница 134: ...8 MM156 DL 1 6 2 pH 7 DO 3 8 4 9 5 9 EC5 DL 1 5 2 6 TDS 3 7 4 datalog 134...
Страница 135: ...5 MSDS SDS EMI 20 7 9 20 7 9 pH pH ORP 5 10 5 15 45 15 6 135...
Страница 140: ...11 140...
Страница 141: ...12 IP67 12 E1 E2 250 nA E2 pH E2 30 141...
Страница 145: ......