![Hach sension 4 Скачать руководство пользователя страница 37](http://html.mh-extra.com/html/hach/sension-4/sension-4_instruction-manual_577114037.webp)
37
SECTION 3,
continued
1. Select the desired channel. From the pH or ISE reading
mode, press the
REVIEW
key. To review the pH calibration, be
sure the meter is in pH mode. To review the ISE calibration,
be sure the meter is in ISE reading mode. View mV values by
pressing the
pH/mV
or
ISE/mV
key.
2. The meter will display the time the calibration was stored.
Press the up arrow key once.
3. The display will show the standard number, standard
concentration and temperature. Press the up arrow once.
4. The meter will continue to scroll through the standard
information with each press of the up arrow key. When all the
standards have been displayed, press the up arrow key again.
5. For pH calibrations, The meter will display the slope and
offset of the calibration curve. Press the up arrow again to see
the r
2
value for the curve. Skip to step 8 if reviewing a
pH calibration.
For ISE calibrations, the meter will display the slope value of
the curve between standards 1 and 2. Press the up arrow once.
6. The meter will display the slope value for the curve between
standards 2 and 3.
7. Each press of the up arrow will display the remaining slope
values until all the slope values have been displayed.
8. To exit Cal Review mode, press
EXIT
. To review any standard
or slope information again, press the down arrow.