38
Operation
4.3.2 Flared tubing connections/switching valve connections
The switching valve is a manual valve designed to direct the flow in the manifold. Some
manifolds that include a column (e.g., cadmium column for nitrate reduction) may have
this type of valve.
A flared tubing connection requires flared tubing 0.5 mm ID from the switching valve flare
kit (Catalog No. 50962). The valve connector is shorter than the tube connector. Refer to
.
To make a connection with the valve:
1.
Slide a metal washer on the flared tubing, then a valve connector. Make sure the
threads of the connector face the metal washer.
2.
Screw the connector into the valve and finger-tighten.
4.3.3 Flow cell connections
Flow cell connections are a combination of the connections described in
To make flow cell connections:
1.
Slide three O-rings over a flow cell flare.
2.
Insert this flare into a flow cell connector so the threaded end faces the O-rings.
Refer to
3.
Screw the connector into the flow cell and finger-tighten.
Figure 38 Connecting flared tubing
1
Tubing
4
Flare
2
Valve connector
5
Switching valve
3
Metal washer
5
1
3
4
2
Figure 39 Flow cell connection
1
Flow cell connector
3
Flare
2
Flow cell
4
O-ring (3x)
1.000
OS
1
2
3
4
Содержание QC8500 Series 2
Страница 2: ......
Страница 6: ...4 Table of Contents ...
Страница 10: ...8 Specifications ...
Страница 30: ...28 Installation ...
Страница 52: ...50 Operation ...
Страница 84: ...82 Troubleshooting ...
Страница 88: ...86 Replacement Parts and Accessories ...
Страница 90: ...88 ...
Страница 92: ...90 Warranty ...
Страница 94: ...92 Certification ...
Страница 102: ...100 Ion Chromatography ...