Installation
3.3.7 Connect the sampler to a PDS200 dilutor
Systems that use a PDS200 dilutor require the ASX-500 Series Sampler and a dual
probe. Refer to
for installation instructions.
3.4 Dilutor installation
The PDS200 uses positive displacement to draw deionized water or other diluent to mix
with non-flammable samples for analysis.
For operation, the PDS200 dilutor requires:
•
An ASX-500 Series autosampler with upgraded firmware, ASX-500 Series serial
number 080313 or greater. For older ASX-500 samplers, contact Technical Support.
•
Installation of Omnion 3.0.218 or higher.
3.4.1 Cable connections
To install the dilutor:
1.
Make sure that the dilutor power switch is off (
).
2.
Turn off and disconnect the power strip.
3.
Connect the Desktop Power Supply with power cable (Catalog No. 5011200) to the
dilutor (
). Plug the other end into the power strip.
4.
Connect the male connector of the 9-pin serial cable to the back of the dilutor
(
). Connect the other end to the dilutor port on the back of the autosampler.
Figure 22 PDS200 dilutor
1
Power indicator light
4
Power cable connector
2
On/Off switch
5
Solenoid valve
3
Serial cable connector
Содержание QC8500 Series 2
Страница 2: ......
Страница 6: ...4 Table of Contents ...
Страница 10: ...8 Specifications ...
Страница 30: ...28 Installation ...
Страница 52: ...50 Operation ...
Страница 84: ...82 Troubleshooting ...
Страница 88: ...86 Replacement Parts and Accessories ...
Страница 90: ...88 ...
Страница 92: ...90 Warranty ...
Страница 94: ...92 Certification ...
Страница 102: ...100 Ion Chromatography ...