12
Installation
3.1.4 Flow cell installation
To install the flow cell:
1.
Insert the flow cell into the detector with the printed side (bottom) down. Since the
flow cell is keyed, insert the flow cell only in the corresponding slot of the detector
(
2.
Make sure the flow cell is pushed all the way in the slot. The bottom flared tubing
connector on the flow cell is the inlet and the top connector is the outlet. During the
installation of a manifold (reaction module), it is very important to make the
connections accordingly, so air bubbles do not get trapped in the flow cell. Refer to
.
3.
Connect the top connector (outlet) to the waste line.
Figure 4 System unit connection
1
On/Off switch
4
Pump connectors
2
Power for system unit
5
USB in (from computer)
3
High pressure pump (IC only)
1
2
1
2
3
5
4
Figure 5 Flow cell
1
To waste
3
Inlet
2
Flow cell body