11
Section 3
Installation
DANGER
Only qualified personnel should conduct the tasks described in this section of the
manual.
3.1 System unit installation
3.1.1 Carrying instructions
CAUTION
Potential Lifting Hazard. A five-channel system weighs 47 kg (103 lb). Do not
unpack, carry or move the system without proper equipment or sufficient number
of people. Make sure that the system is lifted with leg support and not back
support. People with a history of back problems or cardiovascular problems are
prohibited to lift, move or unpack this system.
•
Lift the System Unit and channels from the bottom. Do not lift by the front panel.
•
Always support the entire instrument (e.g., with a cart or a board) during transport.
3.1.2 Connect the power strip
DANGER
Electrical Shock Hazard. The facility circuit breaker rating must not exceed 20 A.
DANGER
Electrical Shock Hazard. A good low impedance earth ground connection is
required to the equipment cord plug.
DANGER
Potential Electrocution Hazard. If equipment is to be used in wet or potentially wet
locations, always connect to power via an outlet with a Ground Fault interrupt
Circuit (GFIC) or a residual Current Circuit Breaker.
To connect the power strip:
1.
Connect the power strip to a power supply.
2.
Position the power strip on the bench so that other components can be connected.
3.
Leave the power strip off during the installation of other system components.
3.1.3 Connect the system unit
Note:
Do not position the equipment in any way that will prevent access to the On/Off switch. Refer
, item 1.
To connect the system unit:
1.
Connect the power cord to the power outlet on the left side of the System Unit
(
). Connect the other end of the power cord to the power strip.
Leave the power strip switch off.
2.
Locate the 9-pin connector of the cable attached to the reagent pump. Plug this
connector in the outlet for Pump 1 (
, item 4) on the interface
panel, located on the left side of the System Unit.
3.
Connect one USB cable to the outlet labeled USB In. The other end will connect to
the computer.
Содержание QC8500 Series 2
Страница 2: ......
Страница 6: ...4 Table of Contents ...
Страница 10: ...8 Specifications ...
Страница 30: ...28 Installation ...
Страница 52: ...50 Operation ...
Страница 84: ...82 Troubleshooting ...
Страница 88: ...86 Replacement Parts and Accessories ...
Страница 90: ...88 ...
Страница 92: ...90 Warranty ...
Страница 94: ...92 Certification ...
Страница 102: ...100 Ion Chromatography ...