Transmitter 9125 – Conductivity / resistivity measurement
Page 35
Temperature electric calibration / Resistor adjust
This calibration is factory realised.
Used to realise an electric calibration for the Pt100
measurement. 2 resistances of known values should
be connected to the temp + and temp - of the module
measure.
This resistors should have a precision about
0,1%.
Proceed as follows :
x
Launch the calibration when you have configured
an electric calibration. Lancer l'exécution après
avoir configuré l'étalonnage en étalonnage
électrique.
x
The transmitter requires the connecting of the first
resistance.
x
Press
Enter.
x
Repeat the same procedure for the second
resistance.
Process calibration
Proceed as follows after you have configured a
process calibration :
x
Wait till the measurement is stable and press the
right function key
OK.
x
You have the possibility to change the value
x
Press
Enter.
x
The instrument executes a zero adjustment
required to display the value configured.
RESISTANCE 1
Press
ENTER
TEMP. CALIB.
189.9
189.9°C
TEMP. CALIB.
1
TYPE :
RES1 : 100
:
TEMP. CALIB
Elect.