5.
Connect the other end of the Tygon
®
tubing to the connector marked
OUT
on the
pump box.
6.
Remount the transparent cell to its cover.
7.
Connect Tygon
®
tubing (4) between the connector marked
IN
on the pump box and
the canister of cleaning solution.
8.
Connect the cable (5) to terminals 4 and 5 of J14 in the electronic unit (refer also to
Figure 8
on page 18 and
Table 6
on page 20).
Note: Make sure that the Tygon
®
tubing (1) does not come into contact with the sample in the
measurement cell.
Reconnect power to the analyzer and proceed as follows:
1.
From the
PROGRAMMING
menu select
SERVICE
.
2.
Select the
CLEANING
option and select
ON
to start it.
3.
Check the pump functions correctly and then select
OFF
to stop it.
4.
From the
PROGRAMMING
menu select
CLEANING
and set the parameters as
required (see
Chemical cleaning (CLEANING option)
on page 30).
8.3 Conditioning
D A N G E R
Electrocution hazard. Always make sure the analyzer is turned off and power is disconnected
before installing this option.
This option allows for the addition of one or two reagents to prepare the sample before
measurement. The kit includes one or two reagent pumps (peristaltic or micro piston) and
one or two 10 liter canisters equipped with level detector and strainer.
To install the pump(s), refer to
Reagent pumps
on page 45.
To install the 10 liter canister(s), refer to
Canister kit
on page 48.
Reconnect power to the analyzer and proceed as follows:
1.
From the
PROGRAMMING
menu select
SERVICE
.
2.
Select the
PUMP 2
option and select
ON
to start it.
3.
Check the pump functions correctly and then select
OFF
to stop it.
4.
Select the
PUMP 3
(if installed) option and select
ON
to start it.
5.
Check the pump functions correctly and then select
OFF
to stop it.
6.
From the
PROGRAMMING
menu select
SAMPLE COND
and set the parameters as
required (see
Sample conditioning (SAMPLE COND option)
on page 31).
8.4 Dilution or decantation
D A N G E R
Electrocution hazard. Always make sure the analyzer is turned off and power is disconnected
before installing this option.
This option is used for the measurement of high concentration samples. The sample is
diluted using water from the rinser, which must be free of any compound being analyzed
in order to avoid any interference. The kit includes one dilution assembly and one rinsing
assembly.
Additional options
40
Содержание POLYMETRON 8810
Страница 1: ...DOC024 52 93057 POLYMETRON Model 8810 ISE Analyzer User Manual 05 2015 Edition 19 ...
Страница 2: ......
Страница 6: ...1 1 Dimensions Figure 1 Panel dimensions mm inches Specifications 4 ...
Страница 7: ...Figure 2 Cabinet dimensions mm inches Specifications 5 ...
Страница 8: ...Specifications 6 ...
Страница 34: ...Menu overview 32 ...
Страница 38: ...Maintenance and Troubleshooting 36 ...
Страница 57: ......