Maintenance

C A U T I O N

 

Multiple hazards. Only qualified personnel must conduct the tasks described in this
section of the document.

Clean the sensor

Clean the sensor when:

"SENSOR?" shows on the display

Stabilization is slow

Results drift or are not accurate

Calibration failure occurs

1.

Soak the sensor in the applicable cleaning agent. Refer to 

Table 3

.

2.

Rinse or soak the sensor in deionized water for 1 minute.

Table 3  Cleaning agents

Contaminant

Cleaning agent

Time

Grease, oils and fats

Electrode cleaning solution

2 hours maximum

Mineral buildup

10% hydrochloric acid (HCl) solution

5 minutes maximum

Replace the batteries

When the battery icon flashes or the tester will not come on, replace all four batteries.
Refer to 

Install the batteries

 on page 5.

Replace the sensor

Note: Only Pocket Pro+ testers have replaceable sensors.

To replace the sensor, refer to the instructions supplied with the sensor.

Troubleshooting

Message

Possible cause

Solution

SENSOR?

The calibration slope is ±10–15%.

Gently clean the sensor. Refer to 

Clean the sensor

on page 10.

ECAL

Calibration failure. The pH calibration
slope is greater than ±15%.

Gently clean the sensor. Refer to 

Clean the sensor

on page 10. Calibrate again. If calibration failure continues,
replace the tester or sensor as applicable.

"- - - -"
(top line)

The parameter value is not within the
measurement range of the tester. Refer to 

Specifications

 on page 2.

Make sure that the value of the sample is within the
measurement range of the tester. Gently clean the sensor.
Refer to 

Clean the sensor

 on page 10. Calibrate as

necessary.

"- - - -"
(bottom line)

The temperature value is not within the
operating temperature range of the tester
or a temperature sensor failure occurred.
Refer to 

Specifications

 on page 2 for the

operating temperature range.

Make sure that the sample temperature is within the
operating temperature range of the tester. Contact
technical support as necessary.

Note: The tester can still be used if a temperature sensor failure
has occurred, but without automatic temperature compensation.

Battery icon
flashes

The batteries have less than 10% power
remaining.

Replace all four batteries. Refer to 

Install the batteries

on page 5.

10

Содержание Pocket Pro

Страница 1: ...User Manual DOC022 53 80394 Pocket Pro Pro pH 1 ...

Страница 2: ...rer be liable for direct indirect special incidental or consequential damages resulting from any defect or omission in this manual The manufacturer reserves the right to make changes in this manual and the products it describes at any time without notice or obligation Revised editions are found on the manufacturer s website Safety information NOTIC E The manufacturer is not responsible for any dam...

Страница 3: ...uropean electrical equipment users must now return old or end of life equipment to the Producer for disposal at no charge to the user Note For return for recycling please contact the equipment producer or supplier for instructions on how to return end of life equipment producer supplied electrical accessories and all auxiliary items for proper disposal Certification Canadian Radio Interference Cau...

Страница 4: ... be used to reduce interference problems 1 Move the equipment away from the device receiving the interference 2 Reposition the receiving antenna for the device receiving the interference 3 Try combinations of the above Product overview The Pocket Pro pH tester and Pocket Pro pH tester measure the pH of general water samples Refer to Figure 1 These testers are waterproof and float Figure 1 Product ...

Страница 5: ...n can cause the release of explosive gases Be sure that the batteries are of the same approved chemical type and are inserted in the correct orientation Do not mix new and used batteries WA RNING Fire hazard Battery substitution is not permitted Use only alkaline batteries Install the four AAA alkaline batteries in the tester Refer to the illustrated steps in Figure 3 Figure 3 Install the batterie...

Страница 6: ...level Flashes when the battery power is less than 10 Shows when the lock feature is on When the lock feature is on the parameter value on the display does not change Push to set the lock feature to on or off Sensor Refer to Troubleshooting on page 10 Shows when the tester is in calibration mode or when a calibration is due If shows next to the calibration icon the last calibration was not successf...

Страница 7: ... Push and hold until SEt shows on the display to go to the settings menu To exit the settings menu push and hold until End shows on the display When in the settings menu push to scroll through the settings Note Power cannot be set to off while in settings or calibration mode Push to set the lock feature to on or off When the lock feature is on the lock icon shows and the parameter value on the dis...

Страница 8: ...es three times 8 Optional To measure another calibration standard maximum of 3 do steps 4 7 again 9 Push and hold to go to continuous measurement mode END shows on the display Note ECAL shows on the display if the calibration was not successful Refer to Troubleshooting on page 10 10 Rinse the sensor and cap with deionized water and blot dry Custom calibration Pocket Pro only Items to collect One o...

Страница 9: ...s 3 7 9 When done with measurements a Rinse the sensor and cap with deionized water b Put the cap on the tester c Set the power to off Advanced operation Configure the settings 1 Push and hold until SEt shows on the display 2 Push to scroll through the settings The current value of the setting shows on the bottom line Option Description Unit Select the temperature unit that shows on the display Ce...

Страница 10: ...10 15 Gently clean the sensor Refer to Clean the sensor on page 10 ECAL Calibration failure The pH calibration slope is greater than 15 Gently clean the sensor Refer to Clean the sensor on page 10 Calibrate again If calibration failure continues replace the tester or sensor as applicable top line The parameter value is not within the measurement range of the tester Refer to Specifications on page ...

Страница 11: ... regions Contact the appropriate distributor or refer to the company website for contact information Table 4 Replacement parts Description Quantity Item no AAA alkaline batteries 4 pkg 4674300 Lanyard 1 201305 SINGLET 7 00 pH 20 pkg 2770120 pH sensor replacement 1 9532001 Table 5 Accessories Description Quantity Item no SINGLET 4 01 pH 20 pkg 2770020 SINGLET 10 01 pH 20 pkg 2770220 Electrode clean...

Страница 12: ...hach com www hach com HACH LANGE GMBH Willstätterstraße 11 D 40549 Düsseldorf Germany Tel 49 0 2 11 52 88 320 Fax 49 0 2 11 52 88 210 info hach lange de www hach lange de HACH LANGE Sàrl 6 route de Compois 1222 Vésenaz SWITZERLAND Tel 41 22 594 6400 Fax 41 22 594 6499 Hach Company Hach Lange GmbH 2014 All rights reserved 12 2014 Edition 2 ...

Отзывы: