1—49
Phosphonates,
continued
Standard Calibration Adjust Method
To perform a standard calibration adjustment using the 2.0 mg/L phosphate
standard (= 20 mg/L due to the dilution), or using an alternative concentration, see
Standard Calibration Adjust on page 2—13
Preparation of 2 mg/L Phosphate Standard Solution
1.
Using a volumetric class A pipet, transfer 2 mL of a 100 mg/L Phosphate
Standard Solution to a 100-mL volumetric flask.
2.
Dilute to the 100-mL mark with deionized water. Stopper and invert to mix.
Interferences
Aluminum
100 mg/L
Arsenate
Interferes at all levels
Benzotriazole
10 mg/L
Bicarbonate
1000 mg/L
Bromide
100 mg/L