• An error has occurred (e.g. sensor out)
3.
When the sample measurement stops because the stop criteria are fulfilled, the gas
concentration and temperature are no longer refreshed. They indicate the measurement when the
stop criteria was reached. If the channel was configured for TPO or TPA calculation, enter the
parameters.
4.
If the sample mode is stopped for another reason (user abort, max. time, or measurement error),
the
aborted
message is displayed.
8.2.5.1 TPO or TPA calculation
The TPO (Total Package Oxygen) and TPA (Total Package Air) features are available on the
instrument with a sensor measuring oxygen. To initiate these options, the instrument must be
configured in
Sample mode
and the TPO or TPA calculation must be enabled. The operator also
needs to ensure the package is shaken for about 5 minutes before measuring, and to know the total
volume of the package and the total volume of the contents of the package.
8.2.5.2 TPO and TPA parameters
1.
Overflow volume: Total package size
2.
Net content volume: Volume of liquid in the package
3.
Press the Compute button to calculate the TPO or TPA. The parameters can be changed and the
value recalculated if necessary.
TPO values are displayed in ppm, TPA values in mL.
4.
To store the measurement, press the
OK
button.
8.2.5.3 Stop criteria configuration
The stop criteria setting is available in the
Sample
mode measurement option in the channel
configuration window.
Note: The parameters available for configuration depend on the type of stop criteria being defined.
1.
Select
Menu>Main>Configure the channel
, then press the
Sample Mode
button and the
Stop
criteria
button.
Option
Description
Above
threshold
The stop criteria is met when the gas concentration is greater than the parameter entered
in
Threshold
Below
threshold
The stop criteria is met when the gas concentration is lower than the parameter entered in
Threshold
Stability
The stop criteria is met when the variation of the gas concentration is smaller than the
parameter entered in
Variation
Note: The number of samples taken into account to compute the Variation can be
adjusted using the parameter Depth.
Time
The stop criteria is met when the elapsed time reaches the
Max. time
parameter.
Max. time
The parameter
Max. time
is the maximum time allowed to reach the target. If
Time
is not
the type of stop criteria, then when this delay has elapsed, the measurement stops and the
message
aborted
is displayed.
Time filter
The time filter allows you to filter the stop criteria. The sample mode is stopped when the
stop criteria is fulfilled for a time greater than the parameter
Time Filter
.
As an example, if criteria is set to
Above threshold
and
Time Filter
is set to 10 sec. the
measurement stops when the gas concentration is above the threshold for more than
10 seconds.
English
19
Содержание ORBISPHERE K-M1100
Страница 278: ...2 3 3 1 3 2 278...
Страница 279: ...3 3 ESD 3 4 2000 m 6562 ft 2000 m 4 GFCI GFI 279...
Страница 280: ...PE 100 240 VAC 5 VDC PE PE 4 1 1 U 2 3 2 4 4 2 1 U 2 3 4 2 5 4 3 280...
Страница 281: ...1 3 4 5 6 7 1 2 3 6 4 4 5 4 1 4 6 7 281...
Страница 282: ...4 4 1 1 6 2 Ethernet 7 3 USB A 8 2 4 USB B 4 9 5 1 4 5 NEMA IP NEMA 4X IP66 4 5 1 1 Ethernet 2 25 mm 8 mm 3 4 282...
Страница 286: ...4 7 2 6 7 J7 J7 16 J7 7 1 2 1 3 2 4 3 5 6 1 7 2 8 3 9 J7 9 J7 12 10 11 12 13 16 4 7 3 NO NC NC NO J14 1 J18 2 J19 3 286...
Страница 287: ...4 8 4 8 1 3 5 10 15 20 15 2 Hach 4 8 2 1100 1100 150 mL min No 3 8 8 1 3 2 4 5 5 1 LED 287...
Страница 288: ...10 1 5 2 5 3 menu Up Main Close Help 5 4 Enter 6 6 1 4 1 Configuration Security ID 288...
Страница 300: ...10 3 10 4 11 RS485 PROFIBUS DP USB HTTP TCP IP 12 12 1 Hach 12 2 6 CIP ClO2 12 2 1 Ksv Ksv 298 1 2 3 300...
Страница 301: ...12 2 2 1 2 3 12 2 3 1 2 3 4 301...
Страница 329: ...M1100 12 PG 13 5 x 246 x 47 0 6 9 69 x 1 85 1 32 K1100 M1100 28 x 143 50 x 49 0 74 5 65 x 1 93 1 63 0 7 2 3 3 1 3 2 329...
Страница 330: ...3 3 3 4 2000 6562 2000 4 GFCI GFI 330...
Страница 331: ...PE 100 240 5 4 1 1 U 2 3 4 4 2 1 U 2 3 4 5 4 3 331...
Страница 332: ...1 3 4 5 6 7 1 2 3 6 4 4 5 4 1 4 6 7 332...
Страница 333: ...4 4 1 1 6 2 Ethernet 7 3 USB A 8 2 4 4 USB B 9 5 1 4 5 NEMA IP NEMA 4X IP66 4 5 1 1 Ethernet 2 25 8 3 4 333...
Страница 334: ...1 4 7 2 5 8 3 6 5 6 4 6 4 6 1 10 30 8 BINDER 2 BINDER 1 10 30 2 3 4 5 6 10 30 7 10 30 8 3 4 6 2 334...
Страница 337: ...4 7 2 6 7 J7 J7 16 J7 7 1 2 1 3 2 4 3 5 6 1 7 2 8 3 9 J7 9 J7 12 10 11 12 13 16 4 7 3 J14 1 J18 2 J19 3 337...
Страница 338: ...4 8 4 8 1 3 5 10 15 20 15 CO2 Hach 4 8 2 150 K1100 M1100 3 8 8 1 3 2 4 5 5 1 338...
Страница 339: ...10 5 2 5 3 Menu Up Main Close Help 5 4 CAP Enter 6 Security 339...
Страница 353: ...12 12 1 Hach 12 2 6 CIP ClO2 12 2 1 Ksv Ksv 350 1 2 3 12 2 2 1 2 3 353...
Страница 354: ...12 2 3 1 2 3 4 354...
Страница 406: ...K1100 M1100 28 mm x 143 50 x 49 mm 0 74 kg 5 65 x 1 93 in 1 63 lbs 0 7 kg 2 3 3 1 3 2 406...
Страница 407: ...3 3 3 4 2000 m 6562 ft 2000 m 4 DC AC GFCI GFI 407...
Страница 408: ...100 240 V AC 5 V DC PE 4 1 1 U 2 3 2 4 4 2 1 U 2 3 4 2 5 4 3 408...
Страница 409: ...1 3 4 5 6 7 1 2 3 4 4 2 5 4 1 4 6 7 409...
Страница 410: ...4 4 1 1 6 2 Ethernet 7 3 USB A 8 2 4 USB B 4 9 5 1 4 5 4X IP66 4 5 1 EMC 1 ethernet 2 25 mm 8 mm 3 4 410...
Страница 411: ...1 4 7 2 5 8 3 6 5 6 4 6 4 6 1 10 30 V DC 8 2 1 10 30 V DC 2 3 4 5 6 10 30 V DC 7 10 30 V DC 8 3 4 6 2 100 240 V AC 4 411...
Страница 415: ...4 8 4 8 1 3 5 10 15 20 15 CO2 Hach 4 8 2 K1100 M1100 150 mL min 3 8 8 1 3 2 4 5 5 1 LED 415...
Страница 416: ...10 5 2 service 5 3 menu Up Main Close Help 5 4 CAP Enter 6 416...
Страница 430: ...12 2 2 1 2 3 12 2 3 1 2 3 4 430...
Страница 531: ......