English
6
7.
Start the measurement by pressing
Read Carousel 1
.
The first carousel insert is measured and the results displayed.
8.
Remove the carousel insert or cells and load carousel insert 2 with
samples.
9.
Start the measurement by pressing
Read Carousel 2
.
The second carousel insert is measured and the results displayed.
Continue until all samples have been measured.
Note: If Instrument Setup > Data Log Setup > Auto Store: On has
been selected, all measurement data is automatically saved. The
measurement data can be accessed under Recall data > Data Log.
Import an external sample ID
1.
Create an external sample ID list with a spreadsheet program.
Four columns are permitted; column 1 contains the sequential
number, column 2 the sample designation, columns 3 and 4 are
optional.
Headings and comment lines must begin with #.
2.
Create a folder named "SampleID" on a USB stick or in your
established network.
3.
Save the sample ID list in CSV or UNICODE-TXT format in the
"SampleID" folder with
Save as
.
4.
Connect the DR6000 to the USB stick or network.
All *.TXT and *.CSV files in the SampleID folder will be displayed for
selection.
5.
Select the desired file with
OK
.
6.
Transfer the displayed sample ID list with
Done
.
Single wavelength
1.
Press
Single Wavelength
in the main menu.
2.
Press
Options > More > Reading mode
.
3.
Select
Carousel 1 inch square
or
Carousel 10 mm square
.
4.
In the toolbar, select the symbol for the carousel insert. This symbol
opens a window with carousel options; refer to
Table 2 Carousel options
Options
Carousel 1 inch
square
Options
Carousel 10 mm
square
Description
Zero 1 Read 1–5
Zero 1 Read 1–7
Position 1 zero solution, from position 1 to
5 (or 7) samples
Measurement stops after the zero
measurement. Afterwards, the carousel
insert is measured one a work step.
Zero 1–5
Read 1–5
Zero 1–7
Read 1–7
Position 1 to 5 (or 7) zero solutions,
position 1 to 5 (or 7) samples.
Measurement stops after all zero
measurements. Afterwards, the samples
are measured in a further work step.
Zero 1 Read 2–5
Zero 1 Read 2–7
Position 1 zero solution, from position 2 to
5 (or 7) samples.
Carousel insert is measured in one work
step.
Zero 1, 3
Read 2, 4
Zero 1, 3, 5
Read 2, 4, 6
Always alternating. First zero solution,
then sample.
Carousel insert is measured in one work
step.
Содержание LZV902
Страница 2: ...English 3 Espa ol 9 Fran ais 17 25 33 Portugu s 41 49...
Страница 16: ...Espa ol 16...
Страница 24: ...Fran ais 24...
Страница 25: ...25 2005 8 12 EU 2002 96 EC EU...
Страница 26: ...26 2 10 mm 7 10 mm 1 5 1 1 2 3 4 1...
Страница 30: ...30 1 2 3 Store Symbol 20 1 4 5 1 2 More Reading mode 3 1 10 mm 4 2 5 6 15 7 1 2 3 Store Symbol 20 1 4 5...
Страница 32: ...32...
Страница 33: ...33 2005 8 12 EU 2002 96 EC EU...
Страница 34: ...34 2 10 mm 7 10 mm 1 5 1 1 2 3 4 1...
Страница 40: ...40...
Страница 48: ...Portugu s 48...
Страница 49: ...49 2005 8 12 2002 96 EC...
Страница 50: ...50 10 7 10 1 5 1 1 2 3 4 1...