4.10.3. Resistivity measurementst:
Resistivity measurement with printer, See EC measurement with printer, page 12.
4. Operation
4.10.4. NaCl, TDS and Resistivity calibration
Standard recognition
The instrument recognises the same standard as in the EC cali-
bration, molar standards.
Calibration to X value
Manual readjustment of the selected unit, NaCl or TDS.
Cell constant introduction
Manual introduction of the cell constant. For example, when the
cell has been calibrated previously in calibration laboratory.
Theoretical calibration
Selecting theoretical calibration the calibration data in memory
will be replaced by C=1.000 cm
-1
.
This is a previous step of the obtention of calibration certificate.
Calibration frequency
Select between 0 h and 7 days. The instrument by default is
programmed for daily calibration and indicates the remaining
time to the new calibration. Selecting 0 h the automatic re-cali-
bration warning is deactivated.