
English
76
To use the 10 mm square cuvette and 11 mm round cuvettes, adapter Z
must be inserted into the cuvette compartment (2). For measurements
with 50 mm square cuvettes, you must remove the adapter.
Take a color measurement
1.
Press
Color Measurement
.
2.
Insert a cuvette/sample cell with distilled water to calibrate.
Note:
Always carry out the calibration very carefully, as a faulty
calibration can cause inaccurate results to be obtained.
3.
The calibration starts automatically once the instrument has
detected the cuvette.
The type of cuvette/sample cell used and the exact progress of
calibration is shown in a separate window.
4.
After calibration, the cuvette size used is displayed in the top right.
Note:
After calibration, you can measure the cuvette with distilled
water again as a sample.
The displayed measured values should match the uncolored color
indices (i.e. Hazen = 0, Gardner = 0.0, CIE L*a*b* = 100.0, 0.0, 0.0,
etc.)
Содержание LICO 620
Страница 3: ...Türkçe 354 英语 374 ...
Страница 4: ......
Страница 24: ...Čeština 23 ...
Страница 84: ...English 83 ...
Страница 106: ...Español 105 ...
Страница 128: ...Français 127 ...
Страница 208: ...한국어 207 ...
Страница 230: ...Nederlands 229 ...
Страница 272: ...Română 271 ...
Страница 314: ...Slovenčina 313 ...
Страница 374: ...Türkçe 373 ...
Страница 378: ...英语 377 正面和背面视图 图 1 正面和背面视图 1 USB 端口 A 类 5 以太网端口 2 试管隔室盖 6 USB 端口 B 类 3 触摸屏 7 USB 端口 A 类 4 开关 8 台式电源连接 ...
Страница 392: ...英语 391 ...
Страница 393: ......