Section 8 Troubleshooting

Problem

Possible cause

Solution

Decreased probe performance
causes slow stabilization and
prevents accurate calibrations or
measurements.

The sensor cap is loose or
damaged.

Tighten or replace the sensor cap. Always
replace the sensor cap and the iButton at
the same time. Refer to 

Replace the

sensor cap and iButton

 on page 9.

Water is between the sensor
cap and the probe lens.

Remove the sensor cap and dry the probe
lens. Refer to 

Clean the probe

 on page 7.

The sensor cap is not
sufficiently conditioned.

Keep the probe in the BOD bottle with
some water for more time, then try the
calibration again.

The temperature or pressure
sensor does not operate
correctly.

Compare the temperature and pressure
readings of the probe with external
measurements. The pressure sensor
reads absolute pressure, which is not
adjusted to sea level. If the measurements
are not correct, contact technical support.

The lot code on the iButton is
not the same as the lot code
on the sensor cap.

Replace the sensor cap and iButton.

Sample properties cause slow
stabilization or inaccurate
measurements.

The measurement was not
adjusted for salinity in the
sample.

Measure the salinity of the sample and
enter the value as a salinity correction
factor in the meter.

Procedure problem causes slow
stabilization and prevents accurate
calibrations or measurements.

Air bubbles are around or
below the probe tip.

Carefully tap or shake the probe to remove
air bubbles.

The LDO sensor cap is in
direct sunlight.

Install the protective shroud.

The calibration mode in the
meter is set to factory.

If user calibration is necessary, change the
calibration mode in the meter.

Section 9 Consumables

Note: Product and Article numbers may vary for some selling regions. Contact the appropriate distributor or refer to
the company website for contact information.

Description

Quantity

Item no.

LDO sensor cap replacement kit (includes iButton)

5811200 

BOD bottle with glass stopper, 300 mL

62100 

Beaker, 250 mL, polypropylene

108046 

Probe stand for standard Intellical probes

8508850 

Probe cable depth markers for rugged Intellical probes

5/pkg

5828610 

Disposable wipes, 11 x 22 cm

280/pkg

2097000 

Wash bottle, polyethylene, 500 mL

62011 

Shroud kit for standard LDO probes

5832500 

Shroud kit for rugged probes

5825900 

12

   

English

Содержание LDO10101

Страница 1: ...DOC022 53 80021 LDO101 05 2022 Edition 5 User Manual ...

Страница 2: ......

Страница 3: ... Preparation for use 5 Section 5 Calibration 6 5 1 Calibration notes 6 5 2 LDO water saturated air 100 calibration procedure 6 Section 6 Sample measurement 7 6 1 Sample measurement notes 7 6 2 Sample measurement procedure 7 Section 7 Maintenance 7 7 1 Clean the probe 7 7 2 Replace the sensor cap and iButton 9 7 3 Remove or install the shroud 9 7 4 Storage 11 Section 8 Troubleshooting 12 Section 9 ...

Страница 4: ...Table of Contents 2 ...

Страница 5: ...ology and a stirrer for measurements in a BOD bottle Figure 1 Probe overview 1 LDO sensor cap 5 Pressure sensor module 2 Rugged probe 6 iButton 1 2 3 Standard probe 7 iButton compartment 4 Cable 8 Pressure sensor module cap Section 2 Specifications Specifications are subject to change without notice Specifications Details Probe type Luminescent dissolved oxygen LDO probe Dissolved oxygen range 0 1...

Страница 6: ...ble connection M12 digital output and connector Warranty 3 years on the probe 1 year on the sensor cap This warranty covers manufacturing defects but not improper use or wear Certifications CE FCC ISED Section 3 Safety information 3 1 Intended use The Intellical probes are intended for use by individuals who measure water quality parameters in the laboratory or in the field The Intellical probes d...

Страница 7: ...are the probe for calibration and measurement as follows Do not touch the protective black layer on the LDO sensor cap 1 Remove the shroud Refer to Remove or install the shroud on page 9 2 Install the LDO sensor cap and tighten by hand Do not tighten more than is necessary for a good seal 3 Soak the probe in 100 mL of tap water for 30 minutes before use For probes that will be in aqueous solutions...

Страница 8: ...bration reminders The calibration data is stored in the probe When a calibrated probe is connected to a different meter with the same calibration options a new calibration is not necessary Air bubbles below the sensor when in solution can cause a slow response or error in the calibration Make sure to remove air bubbles during calibration The meter uses the slope value shown at the end of the calib...

Страница 9: ...before and during measurements Make sure to install the shroud before field use to prevent damage to the sensing elements Refer to Remove or install the shroud on page 9 The probe warranty does not include such damage To deploy a rugged probe at a distance toss the probe body with a slow underhand throw Do not throw the probe by the cable to prevent damage to cable the probe or the user 6 2 Sample...

Страница 10: ...asive cleaners c Install the LDO sensor cap and tighten by hand Do not tighten more than is necessary for a good seal 3 Put the probe in a neutral cleaning solution and stir the solution Do not rub or remove the black layer on the LDO sensor cap Note If disinfection is necessary put the probe in a dilute alcohol solution for a maximum of 2 minutes and rinse immediately with deionized water Do not ...

Страница 11: ...5 days or more frequently if the cap becomes damaged or dirty The LDO sensor cap and iButton operate together and must be replaced at the same time Refer to the instructions that are supplied with the LDO sensor cap replacement kit 7 3 Remove or install the shroud Remove the shroud during calibration and maintenance Refer to Figure 3 Keep the shroud installed on the probes during sample measuremen...

Страница 12: ...Figure 3 Remove the shroud 10 English ...

Страница 13: ...aily measurements in the laboratory keep the probe in a humid environment such as a BOD bottle filled approximately full with tap water If the probe is used to monitor DO concentrations continuously for more than 6 hours keep the probe in tap water When the probe is used for short periods less than 6 hours the storage can be dry Before dry storage rinse the probe in deionized water and blot dry So...

Страница 14: ...slow stabilization or inaccurate measurements The measurement was not adjusted for salinity in the sample Measure the salinity of the sample and enter the value as a salinity correction factor in the meter Procedure problem causes slow stabilization and prevents accurate calibrations or measurements Air bubbles are around or below the probe tip Carefully tap or shake the probe to remove air bubble...

Страница 15: ......

Страница 16: ...rs hach com www hach com HACH LANGE GMBH Willstätterstraße 11 D 40549 Düsseldorf Germany Tel 49 0 2 11 52 88 320 Fax 49 0 2 11 52 88 210 info de hach com www de hach com HACH LANGE Sàrl 6 route de Compois 1222 Vésenaz SWITZERLAND Tel 41 22 594 6400 Fax 41 22 594 6499 Hach Company Hach Lange GmbH 2010 2013 2021 2022 All rights reserved ...

Отзывы: