Table 7 Sensor warning list (continued)
Warning
Definition
Resolution
Temperature < 0 °C
The process temperature is below 0 °C
(32 °F)
Increase the process temperature or
stop use until the process temperature
is in the sensor specification range.
Temperature > 50 °C
The process temperature is above 50 °C
(120 °F)
Decrease the process temperature or
stop use until the process temperature
is in the sensor specification range.
Red amplitude is too low.
Value falls below 0.03
on page 9 .
Red amplitude is too high.
Value is greater than 0.35
Call technical support.
Blue amplitude is too low.
Value is below 0.03
on page 9.
Blue amplitude is too high.
Value is greater than 0.35
Call technical support.
Sensor cap lot code
missing.
The sensor cap code has become
corrupt. The code has been reset
automatically to the default cap and lot
codes.
Complete the sensor setup cap
procedure. If no setup cap is available
for the sensor cap, call technical
support.
Event list
The Event list keeps a log of changes to how data is recorded by the sensor. To show sensor events,
go to MENU>SENSOR DIAGNOSTICS>[Select Sensor]>EVENT LIST.
Refer to
Table 8 Event list for the sensor
Event
Description
Altitude/Pressure unit change
Atmospheric pressure or altitude units have changed.
Altitude/Pressure change
The value for altitude or atmospheric pressure has changed.
Temperature change
The units for temperature have changed.
Unit change
A new unit of measurement has changed.
Salinity change
The value for salinity has changed.
Reset
Sensor settings have been reset to the default values.
Sensor setup change
The sensor setup has changed.
Cleaning interval change
The time between sensor cleaning has changed.
Sensor life change
The time between sensor cap replacements has changed.