Commissioning

Prepare the LOC 100 RFID read/write device

1.

Open the battery compartment by releasing the screw on the back 
side.

2.

Insert two type AA batteries into the battery compartment; note the 
poles.

3.

Close the battery compartment and secure it with the screw.

4.

Switch on the LOC 100 using the On/Off key.

a.

To charge the batteries, connect the LOC 100 to the 
photometer via a USB cable. Alternatively, the batteries can be 
charged via a USB cable using a standard PC or laptop.

b.

It is also possible to charge the batteries in a standard external 
charger for NiMH rechargeable batteries.

c.

The battery charge is shown via a status display on the 
LOC 100 screen. 
The nominal capacity of the batteries is not reached until 
repeated charge/discharge cycles have been completed. 

5.

Press and hold down the On/Off key to open the Configuration 
menu. Use the three keys to adjust the brightness, contrast, date, 
time, tone and reader of the operator ID.

Note: If the LOC 100 is connected to the photometer via a USB 
cable, the rechargeable batteries are charged and the date and time 
are transferred from the photometer to the LOC 100. 

Note: All settings can also be made from the PC. To do so, connect the 
LOC 100 to the PC using a USB cable. The LOC 100 is displayed as an 
RFID mass storage device; the settings for brightness, contrast, date, 
time and tone can be changed in the RFID.exe. file.

D A N G E R

Risk of fire and explosion. 
Only use the NiMH batteries provided in the measuring instrument; no other 
battery types are permitted.

Make sure that the batteries are inserted correctly into the battery compartment. 
Incorrect insertion of the batteries could cause damage to the instrument, fire or 
explosions.

Содержание LANGE LOC 100

Страница 1: ...DOC012 97 90327 LOC 100 06 2011 Edition 2 Basic User Manual en Manuel d utilisation de base fr Manual do utilizador b sico pt...

Страница 2: ...English 3 Fran ais 15 Portugu s 27...

Страница 3: ...x 64 pixels Input 3 touch keys Power supply 2 rechargeable AA Ni MH batteries included within the scope of delivery Operating hours 1000 work cycles approx 2 months Charging time required for full cha...

Страница 4: ...zation in other countries In case of doubt please contact the distributor The LOC 100 contains an RFID module to receive and transmit information and data The RFID module operates with a frequency of...

Страница 5: ...nce with the instruction manual may cause harmful interference to radio communications Operation of this equipment in a residential area is likely to cause harmful interference in which case the user...

Страница 6: ...hotometer so that it can be saved together with the test results This allows the sample location sample time and the operator to be documented consistently The RFID starter set for sample identificati...

Страница 7: ...English 7 Figure 1 Front view of LOC 100 1 Left key Read 4 Mini USB port 2 On Off key 5 Display 3 Right key Write...

Страница 8: ...English 8 Figure 2 Rear view of LOC 100 1 Screw 3 RFID module 2 Battery compartment cover 4 AA rechargeable battery 2x...

Страница 9: ...discharge cycles have been completed 5 Press and hold down the On Off key to open the Configuration menu Use the three keys to adjust the brightness contrast date time tone and reader of the operator...

Страница 10: ...or ID refer to photometer operating instructions 2 Select Initialize RFID Tag 3 Follow the onscreen instructions and hold the operator RFID tag in front of the photometer RFID module A sound confirms...

Страница 11: ...tep is optional The option to read out the operator ID can be deactivated in the Configuration menu 3 Read in the sample ID initialized on the photometer by holding the LOC 100 above the location RFID...

Страница 12: ...n the photometer 2 To determine the sample parameters prepare the sample in accordance with the process specification for the relevant tests Insert the prepared cell into the photometer cell compartme...

Страница 13: ...function was deactivated Check whether the correct tag is being used for the procedure E 04 Data format error The data structure on the tag is incorrect Tag is damaged and must be replaced Replacement...

Страница 14: ...English 14...

Страница 15: ...ixels Entr e 3 boutons tactiles Alimentation 2 piles rechargeables AA Ni MH incluses dans la livraison Heures de fonctionnement 1 000 cycles de fonctionnement env 2 mois Temps n cessaire pour une char...

Страница 16: ...0 contient un module RFID permettant de recevoir et de transmettre des informations et des donn es Le module RFID fonctionne une fr quence de 13 56 MHz DANGER Indique une situation de danger potentiel...

Страница 17: ...dentielle risque de causer des interf rences nuisibles dans ce cas l utilisateur doit corriger les interf rences ses frais Les techniques ci dessous peuvent permettre de r duire les probl mes d interf...

Страница 18: ...c les r sultats de test Cela permet de conna tre exactement l emplacement de l chantillon la date et l heure auxquelles il a t pr lev ainsi que l utilisateur qui a proc d au pr l vement Le kit de d ma...

Страница 19: ...Fran ais 19 Figure 1 Vue avant du LOC 100 1 Bouton gauche Lire 4 Port mini USB 2 Bouton Marche Arr t 5 Ecran 3 Touche droite Ecrire...

Страница 20: ...Fran ais 20 Figure 2 Vue arri re du LOC 100 1 Vis 3 Module RFID 2 Cache du compartiment piles 4 Pile rechargeable AA x 2...

Страница 21: ...la touche Marche Arr t et maintenez la enfonc e pour ouvrir le menu Configuration Utilisez les trois touches pour r gler la luminosit le contraste la date l heure le volume sonore et la lecture de l I...

Страница 22: ...reportez vous aux instructions d utilisation du photom tre 2 S lectionnez Initialize RFID Tag initialiser l tiquette RFID 3 Suivez les instructions l cran et placez l tiquette RFID utilisateur devant...

Страница 23: ...sactiver l option de lecture de l ID de l op rateur dans le menu Configuration 3 Lisez l ID d chantillon initialis avec le photom tre en tenant le LOC 100 au dessus de l tiquette RFID d emplacement e...

Страница 24: ...hotom tre 2 Pour d finir les param tres de l chantillon pr parez l chantillon conform ment aux sp cifications du processus pour les tests correspondants Ins rez la cuve pr par e dans le compartiment p...

Страница 25: ...quette appropri e pour cette proc dure E 04 Erreur de format de donn es La structure des donn es sur l tiquette est incorrecte L tiquette est endommag e et doit tre remplac e Pi ces de rechange Descri...

Страница 26: ...Fran ais 26...

Страница 27: ...tela Fonte de alimenta o 2 pilhas recarreg veis AA Ni MH inclu do dentro do escopo de entrega Horas de opera o Aproximadamente 1 000 ciclos de trabalho 2 meses necess rio tempo de carga completo para...

Страница 28: ...m caso de d vida entre em contato com o distribuidor O LOC 100 cont m um m dulo RFID para receber e transmitir informa es e dados O m dulo RFID opera com uma frequ ncia de 13 56 MHz PERIGO Indica uma...

Страница 29: ...sejada caso em que o usu rio ser solicitado a corrigir a interfer ncia por conta pr pria As seguintes t cnicas podem ser usadas para reduzir problemas de interfer ncia 1 Desconecte o equipamento de su...

Страница 30: ...os do teste Isso permite que a localiza o e o hor rio de coleta da amostra al m do operador que a coletou sejam documentados de maneira consistente O conjunto iniciante de RFID para identifica es de a...

Страница 31: ...Portugu s 31 Figura 1 Vista frontal do LOC 100 1 Tecla da esquerda Leitura 4 Mini porta USB 2 Tecla ligar desligar 5 Tela 3 Tecla da direita Escrita...

Страница 32: ...Portugu s 32 Figura 2 Vista posterior do LOC 100 1 Parafuso 3 M dulo RFID 2 Tampa do compartimento da bateria 4 Pilha recarreg vel AA 2x...

Страница 33: ...amente conclu dos 5 Pressione e segure o bot o Ligar Desligar para abrir o menu de Configura o Use os tr s bot es para ajustar o brilho o contraste a data a hora a tonalidade e o leitor do ID do opera...

Страница 34: ...es de opera o do fot metro 2 Selecione Inicializar R tulo de RFID 3 Siga as instru es na tela e segure a etiqueta RFID do operador na frente do m dulo RFID do fot metro Quando o processo for conclu do...

Страница 35: ...o ID do operador pode ser desativada no menu de Configura o 3 Para fazer a leitura do ID de amostra inicializado no fot metro segure o LOC 100 acima da etiqueta RFID de localiza o e pressione rapidame...

Страница 36: ...metro 2 Para determinar os par metros da amostra prepare a amostra de acordo com as especifica es do processo dos testes relevantes Insira a c lula preparada no compartimento da c lula do fot metro 3...

Страница 37: ...st sendo utilizado para este procedimento E 04 Erro de formato de dados A estrutura de dados no r tulo est incorreta O r tulo est danificado e precisa ser substitu do Pe as de reposi o Descri o Cat no...

Страница 38: ...Portugu s 38...

Страница 39: ...www hach com www hach lange com HACH Company 2011 All rights reserved Printed in Germany...

Отзывы: