Select Single Wavelength options
1.
Push
SETTINGS>Photometer Functions>Single Wavelength>Options>Advanced Options
.
2.
Select the applicable options.
Option
Description
λ
Select a wavelength between 340 to 800 nm. Default: 560 nm
Concentration
Select the applicable unit and enter the factor. Defaults: mg/L and 1.000
Resolution
Select the resolution with the necessary number of decimal places. Default: 0.01
Save to User Programs
Save the selected parameter as an unique user program. Select the name, unit,
wavelength, resolution, chemical formulas 1–4, calibration formula, upper and
lower limit and the timer 1–4.
Select Multi Wavelength options
1.
Push
SETTINGS>Photometer Functions>Multi Wavelength>Options>Advanced
Options>λ1–λ4
.
2.
Select the applicable options.
Option
Description
Wavelength
Select two or more wavelengths. Defaults: 400, 500, 700 or 800 nm.
Abs Formula
Select the applicable absorbance formulas for the calculation of the multi-wavelength reading.
The formula specifies the wavelength and the coefficients.
Factors
Select the multiplication factors for converting absorbance values into concentration values.
Select Time Course options
1.
Push
SETTINGS>Photometer Functions>Time Course>Options
.
2.
Select the applicable options.
Option
Description
Total Minutes
Enter the total time in minutes for the Time Course. Range: 1 to 60. Default: 10 minutes
Interval Seconds
Enter the interval in seconds for the Time Course. Range: 10 to 600. Default: 30 seconds
Wavelength
Select a wavelength between 340 and 800 nm. Default: 560 nm
Data Recall
Show the saved Time Course data.
Show Table
Show the saved Time Course data in a table.
Exit
Exit the current menu.
Use an operator ID
The operator ID tag associates readings with an individual operator. All stored data will include this
ID.
1.
Push
Options>Operator ID
in the reading screen.
2.
Select, create or delete an operator ID:
Option
Description
Current ID
Select an ID from a list. The current ID will be associated with the sample data