9
Section 3
Installation
DANGER
Electrocution and fire hazard. Only qualified personnel should conduct the tasks
described in this section of the manual.
3.1 Unpack the instrument
Remove the analyzer from its shipping carton and inspect it for any damage. Verify that
the Installation Kit, Cat. No. 5516400, and Maintenance Kit, Cat. No. 5444300, are
included aloing with the reagents.
If any items are missing or damaged, contact the manufacturer or distrubutor (outside the
US).
3.2
Instrument environmental considerations
The instrument enclosure is designed for general-duty, indoor installation with a nearly
constant ambient temperature between 5° and 40 °C (18° to 104 °F). The enclosure
environmental rating is IP62 with the door closed and latched. Do not mount the
instrument in direct sunlight; shield from dripping water.
3.3 Mount the instrument on the wall
The instrument case is designed for wall mounting. Refer to
Figure 2
,
Figure 3
, and
Figure 4
for dimensions and other installation information. Use ¼-inch screws for
mounting. Mount the instrument as close to the sampling point as practical to ensure
complete purging of the sample line during each cycle. Leave adequate clearance at the
sides and bottom of the instrument case for plumbing and wiring connections.
Содержание Lange CL17
Страница 2: ......
Страница 13: ...11 Installation Figure 3 CL17 instrument dimensions 2 of 3 ...
Страница 18: ...16 Installation Figure 7 Sample conditioning kit ...
Страница 32: ...30 System Startup ...
Страница 40: ...38 Operation ...
Страница 50: ...48 Troubleshooting ...
Страница 52: ...50 Parts and accessories ...
Страница 56: ...54 Certification ...
Страница 60: ...58 Network Interface Card ...