SECTION 5,
continued
61
Figure 15 illustrates a suggested hookup using a pressure regulator in the inlet
line and a flow meter or flow control valve on the outlet line. The control
valve is installed on the outlet side because it may introduce bubbles into the
sample causing positive interference in the turbidity reading. This hookup also
maintains sufficient pressure to minimize outgassing in a carbonated stream.
Figure 14
Compression Fitting
Figure 15
Suggested High Pressure Installation
Nut
o-ring
Spacer
Grab Ring
Tubing
Sample In
Pressure
Regulator
Flow-Through Cell
Flow Meter
(optional)
Return
Or
To Drain
Control
Valve
Содержание Lange 2100AN
Страница 6: ...6...
Страница 8: ...8...
Страница 10: ...10...
Страница 14: ...14...
Страница 46: ...46...
Страница 48: ...48...
Страница 64: ...64...
Страница 74: ...74...
Страница 76: ...76...
Страница 78: ...78...
Страница 84: ...84...
Страница 92: ...92...
Страница 102: ...102...
Страница 103: ...103 HOW TO ORDER...
Страница 106: ...106...
Страница 107: ...107...