Section 8 Maintenance

Regular maintenance is necessary for the best accuracy, stabilization time and life of the probe.
Keep the probe in the recommended storage solution between measurements.

8.1 Clean the probe

Clean the probe regularly to remove contamination and to keep the reference junction open.
Symptoms of contamination:

• Incorrect or irregular readings
• Slow stabilization times
• Calibration errors
• Sample material stays on the probe

1.

Rinse the probe with deionized water. Use warm (35–45 °C (95–113 °F)) deionized water to
remove storage solution that dries on the probe. Dry the probe body with a lint-free cloth.

2.

Soak the glass bulb and reference junction in the applicable cleaning solution for the specified
time. Refer to 

Table 1

 and 

Consumables

 on page 15.

3.

Rinse or soak the probe for 1 minute in deionized water. Dry the probe body with a lint-free cloth.

4.

Soak the probe in pH 4 buffer for 20 minutes.

5.

Rinse the probe with deionized water. Dry the probe body with a lint-free cloth.

Table 1  Cleaning solution

Contamination

Cleaning solution

Active component

Soak time

General
contamination

Electrode cleaning solution
for regular maintenance

KATHON

 CG,

DECONEX

®

11 

12–16 hours

Minerals

Electrode cleaning solution
for minerals/inorganic
contamination

Phosphoric acid
(~10%)

10–15 minutes

Fats, grease and oils

Electrode cleaning solution
for fats, oils and grease
contamination

KATHON

 CG,

TRITON

®

 X

2 hours maximum

Proteins

Electrode cleaning solution
for proteins/organic
contamination

Pepsin in HCl

3 hours maximum

Wastewater and
organic compounds

Electrode cleaning solution,
extra strong

Sodium hypochlorite

5–10 minutes

Blockage of porous
pin/diaphragm
reference junctions

Electrode cleaning solution
for porous pin/diaphragm
reference junctions

Thiourea

3 hours with stirring
or until the porous
plug becomes white

English

   

11

Содержание Intellical PHC805

Страница 1: ...DOC022 53 80436 PHC805 05 2021 Edition 2 User Manual...

Страница 2: ......

Страница 3: ...7 5 2 Calibration procedure 7 Section 6 Sample measurement 8 6 1 Sample requirements 8 6 2 Sample measurement notes 9 6 3 Sample measurement procedure 9 Section 7 Verify the calibration 9 7 1 Verifica...

Страница 4: ...Table of Contents 2...

Страница 5: ...mples of low ionic strength For best accuracy do not use the probe to measure the pH of very alkaline samples Figure 1 Probe overview 1 Probe soaker bottle with storage solution 5 Probe soaker bottle...

Страница 6: ...acturing defects but not improper use or wear Certifications CE FCC ISED Section 3 Safety information 3 1 Intended use The Intellical probes are intended for use by individuals who measure water quali...

Страница 7: ...ion for use N O T I C E Make sure to remove the protective tape from the filling hole of new probes A probe with a blocked filling hole will not operate correctly New probes are filled with electrolyt...

Страница 8: ...Figure 2 Remove the protective tape Figure 3 Probe position during use 1 Open filling hole 3 Level of calibration solution or sample 2 Level of electrolyte filling solution 6 English...

Страница 9: ...more than one probe is connected to the meter if applicable Calibrate the probes and verify the calibration regularly for best results Use the meter to set calibration reminders The calibration data...

Страница 10: ...e meter documentation for meter operation and probe specific settings 6 1 Sample requirements N O T I C E Make sure that the sample type is compatible with the probe or probe damage can occur Samples...

Страница 11: ...mple 2 Open the filling hole 3 Rinse the probe with deionized water Dry the probe with a lint free cloth 4 Put the probe in the sample with the sensor and reference junction fully in the sample Do not...

Страница 12: ...ionized water Dry the probe with a lint free cloth 5 Put the probe in the pH buffer solution with the sensor and reference junction fully in the solution Do not put the probe on the bottom or sides of...

Страница 13: ...ater Dry the probe body with a lint free cloth 4 Soak the probe in pH 4 buffer for 20 minutes 5 Rinse the probe with deionized water Dry the probe body with a lint free cloth Table 1 Cleaning solution...

Страница 14: ...eplace the filling solution 1 Tilt the probe and open the filling hole 2 Use a plastic transfer pipet to remove the contaminated solution from the filling hole Discard the solution 3 Rinse the inner p...

Страница 15: ...s of low ionic strength For short term storage the probe can stay in the sample for a maximum of 2 hours if the sample pH is not high Close the filling hole and put the soaker bottle that contains the...

Страница 16: ...ilization or inaccurate measurements The sample absorbs carbon dioxide CO2 from the air which causes the pH value to slowly decrease in low ionic strength LIS or high purity samples Use the LIS chambe...

Страница 17: ...ckets 20 mL each 20 pkg 2770020 pH 7 00 buffer solution Singlet one use packets 20 mL each 20 pkg 2770120 pH 10 01 buffer solution Singlet one use packets 20 mL each 20 pkg 2770220 pH 4 01 and pH 7 00...

Страница 18: ...escription Quantity Item no Beaker 30 mL plastic colorless 80 pkg SM5010 Beaker 30 mL plastic red 80 pkg SM5011 Beaker 30 mL plastic yellow 80 pkg SM5012 Beaker 30 mL plastic blue 80 pkg SM5013 Beaker...

Страница 19: ......

Страница 20: ...669 2932 orders hach com www hach com HACH LANGE GMBH Willst tterstra e 11 D 40549 D sseldorf Germany Tel 49 0 2 11 52 88 320 Fax 49 0 2 11 52 88 210 info de hach com www de hach com HACH LANGE S rl 6...

Отзывы: