37
SECTION 9 TROUBLESHOOTING for additional information on the system warnings.
To display or clear the WARNINGS, press
SETUP
to call up the setup menu:
Press the
UP ARROW
key to scroll to:
Press
ENTER
.
Press
ENTER
. The system displays the Warning Codes shown in Table 3. Additional details
about the warning can be found in Section 9.1.2 Error Codes and Section 9.1.3 System
Warnings. The warning codes appear momentarily followed by CLeaR WaRNing.
Note: The probe errors
and the reference timer
time-out require the cause
of the fault to be corrected
before they will clear.
Press
ENTER
to CLeaR WaRNings on the display.
2.3
Calibration
Use this menu to calibrate the pH (or ORP) electrode, review the pH calibration, calibrate
the temperature sensor, and adjust the mV offset. The calibration menu is entered by
pressing the CAL key. The first display to appear is STD 1, followed by STD 2 (after STD
1 is completed), GRAB SMP, REVIEW, TEMP, and MV ZERO. The user may then scroll
up or down through the first level of the menu to make adjustments to each calibration.
(Note: If the keyboard is locked, only the REVIEW sub menu may be entered.)
Table 3 Warning Codes
CODE
MESSAGE
CAUSE
E00
EE FAIL
EEPROM failure
E01
REF TIME
reference timer time-out
E02
SOFT SRT
software restart
E03
COLD SRT
cold restart
E04
PROBE ER
pH probe open
E05
PROBE ER
pH probe shorted
E06
PROBE ER
temperature probe open
E07
PROBE ER
temperature probe shorted
E08
MARG CAL
marginal pH offset
E09
MARG CAL
marginal pH gain
E10
MARG CAL
marginal temperature offset
E11
MARG CAL
marginal temperature gain
E12
PWR FAIL
power failure
Содержание EC1000
Страница 2: ...2...
Страница 8: ...8...
Страница 12: ...12...
Страница 14: ...14...
Страница 16: ...16...
Страница 18: ...18 Figure 1 Controller Front Panel...
Страница 20: ...20 Figure 2 Pump Module...
Страница 24: ...24 Figure 6 pH Control Loop...
Страница 54: ...54...
Страница 56: ...56...
Страница 58: ...58 Figure 12 Controller Installation Panel Mount...
Страница 61: ...61 Figure 15 Installing the pH ORP Electrode Figure 16 Flow Thru Cell...
Страница 70: ...70...
Страница 73: ...73 6 4 Calibration See Section 2 3 Calibration Figure 21 Priming Tee Operation...
Страница 74: ...74...
Страница 80: ...80...
Страница 83: ...83 Figure 24 Sensor Replacement...
Страница 85: ...85 Figure 25 Hach One pH Process Electrode...
Страница 89: ...89 Figure 26 Pump Module Plumbing Diagram...
Страница 102: ...102...
Страница 103: ......
Страница 104: ......
Страница 105: ......
Страница 106: ......
Страница 108: ...108...
Страница 112: ...112...
Страница 116: ...116...
Страница 122: ...122...
Страница 123: ...123...
Страница 124: ......