23
1.8 Principle of Operation
The EC1000 Process pH/ORP System consists of a combination pH or ORP electrode, a
controller, and a reference electrolyte pump (see Figure 5 on page 23). The electrode
consists of a pH or ORP measuring half cell and a double junction reference half cell,
which establish a potential characteristic of the sample. This potential is converted to a
proportional current by a built in preamplifier in order to stabilize the signal for long
distance transmission to the controller. Included in the electrode is a temperature sensor
whose signal also is stabilized and sent to the controller to allow for temperature
compensation or temperature correction of the pH/ORP measurement. Separate from the
controller and electrode is a reference electrolyte pump that supplies fresh electrolyte to
the reference half cell. The pumped electrolyte maintains an open reference junction
through a capillary outlet to produce a stable and constant reference potential.
Note: The electrolyte
pump is not used with the
self-pressurized Industrial
Process pH Electrode with
self-pressurized
electrolyte, Cat. No.
48276-00. See Section 1.6
Industrial Process pH
Electrode.
As the electrode measures the pH (or mV) of the water source, a signal is transmitted to
the controller. If the pH (or mV) is outside the setpoint range, the controller can transmit
a current output signal to a reagent pump (or valve) that adds reagent to a mixing tank. See
Figure 6 on page 24 for details. As the reagent is added to the water, the value of the pH
(or redox potential) is adjusted to fall within the setpoint range. Refer to SECTION 4, PID
CONTROL, for a detailed discussion of this process.
Figure 5
System Components
Содержание EC1000
Страница 2: ...2...
Страница 8: ...8...
Страница 12: ...12...
Страница 14: ...14...
Страница 16: ...16...
Страница 18: ...18 Figure 1 Controller Front Panel...
Страница 20: ...20 Figure 2 Pump Module...
Страница 24: ...24 Figure 6 pH Control Loop...
Страница 54: ...54...
Страница 56: ...56...
Страница 58: ...58 Figure 12 Controller Installation Panel Mount...
Страница 61: ...61 Figure 15 Installing the pH ORP Electrode Figure 16 Flow Thru Cell...
Страница 70: ...70...
Страница 73: ...73 6 4 Calibration See Section 2 3 Calibration Figure 21 Priming Tee Operation...
Страница 74: ...74...
Страница 80: ...80...
Страница 83: ...83 Figure 24 Sensor Replacement...
Страница 85: ...85 Figure 25 Hach One pH Process Electrode...
Страница 89: ...89 Figure 26 Pump Module Plumbing Diagram...
Страница 102: ...102...
Страница 103: ......
Страница 104: ......
Страница 105: ......
Страница 106: ......
Страница 108: ...108...
Страница 112: ...112...
Страница 116: ...116...
Страница 122: ...122...
Страница 123: ...123...
Страница 124: ......