Section 5
Repairs
D A N G E R
Electrocution hazard. Remove power from the instrument before this procedure is started.
W A R N I N G
Chemical exposure hazard. Obey laboratory safety procedures and wear all of the
personal protective equipment appropriate to the chemicals that are handled. Refer to the
current safety data sheets (MSDS/SDS) for safety protocols.
5.1 Open the instrument
Items to collect:
•
Coin
•
TORX 10 torque screw driver
•
Krytox oil (item no. 16004801)
Step Description
Picture
1
To open the battery compartment, use a coin.
2
Remove the batteries.
35