Error
Description
User action recommendations
Technical information including troubleshooting action
items (only for Service Operator)
E-21
Signal
measurement
unstable
There is an unstable signal on
the light detector. There is too
much or unstable ambient light.
Make sure that the instrument
cap is fully installed on the cell
holder. Do the test again. If the
error continues, contact technical
support.
Unstable signal on light detector. Too much or unstable
ambient light? Internal frequency measurement 'ratio of
Highfrequency/Midfrequency' out of range.
+ Check/read the service log for possible other error messages.
Check the absorbance reading in 'service mode'. Test
measurment can be performed with any vial...
>
> Replace detector
> Replace LED
E-22
Hardware error
The electronic system is
defective. Contact technical
support.
Mainboard hardware error with ADC system.
ADC Timeout Error or ADC Invalid Status Error (Battery
Voltage, Hardware ID voltage, Module ID voltage, LED monitor
voltage, LED voltage, Temperature)
+ Check/read the service log for possible other error messages.
>
E-30
No application
There was an error during the
application update. A valid
application was not found on the
instrument. Update the
instrument again.
No valid application found on the device.
+ Check/read the service log for possible other error messages.
> Update application to device again.
>
E-31
Bootloader
update failed
There was an error during the
transmission of the bootloader
update. Update the bootloader
again.
Bootloader Update failed during transmission. The device is still
intact.
Error feedback will also be shown on the updating mobile
phone.
+ Check/read the service log for possible other error messages.
> Update bootloader to device again.
>
E-32
Application
update failed
There was an error during the
transmission of the application
update. Update the instrument
again.
Application update failed during transmission. The bootloader is
still intact.
Error feedback will also be shown on the updating mobile
phone.
> Update application to device again.
>
E-66
Update failed
The instrument is defective.
Contact technical support.
Device is not functional anymore.
>
6.1 Get the instrument logger
1.
Install the service cable VAA10036.99.G00 in the DR300. Refer to
2.
Start the VAP235 software.
3.
Connect the service cable VAA10036.99.G00 to an USB port.
4.
Push on the DR300.
5.
Push
Connect
. The display of the DR300 shows serial.
6.
Push
Get instrument log
. Refer to
7.
Find the txt file in the path which is defined in Settings > Program Settings > General
> Instrument log path. E.g. Logger_LPG445_18060A000047.txt. Refer to
Troubleshooting
56
Содержание DR300
Страница 1: ...DOC222 53 90639 DR300 Service Manual 02 2019 Edition 2...
Страница 2: ......
Страница 4: ...Table of Contents 2...
Страница 6: ...Specifications 4...
Страница 15: ...Startup 13...
Страница 16: ...Startup 14...
Страница 26: ...Step Description Picture 10 Check if all segments are displayed and press SELECT button Inspection 24...
Страница 36: ...Inspection 34...
Страница 54: ...Repairs 52...
Страница 65: ...Figure 15 Set delivery state Troubleshooting 63...
Страница 66: ...Troubleshooting 64...
Страница 82: ...Certificate of Service Inspection 80...
Страница 84: ...Certificate of Service Inspection manually 82...
Страница 85: ......