Specifications
5940018Spec.fm
Page 9
Specifications
Wavelength Range
400 to 880 nm
Bandwidth
4 nm ± 1 nm
Wavelength Accuracy
± 1 nm
Wavelength Resolution
1 nm
Wavelength Selection
Automatic, based on method selection
Optical System
Concentric Spectrometer for multi-channel spectroscopy
Wavelength Calibration
Automatic, via internal filter
Linear Photometric Range
–2.000 to +2.000 Abs
Photometric Linearity
±0.005 Abs from 0.0 to 0.5 Abs; ±1% from 0.5 to 2.0 Abs.
Photometric Resolution
0.001 Abs, 0.1 %T
Recommended Recertification Interval
One year (See
.)
Stray Light
>2.5 A, <0.3 %T @ 430 nm using 3 mm thickness GG-475 filter
Operational Modes
Selectable; momentary (Display Lock on), or constant (Display Lock off)
Readout Modes
Transmittance, absorbance, and concentration
RS232 Port
Standard 9-pin bi-directional
Line Power
9 VDC @ 1 amp; 95 to 240 VAC; 50/60 Hz; automatic selection
Battery Power
3 “D” Cell Alkaline or Rechargeable Alkaline Batteries
Sample Compartment and
Cell Compatibility:
Note:
For Hach Programs, be sure to use the cell type specified for that
program in the DR/2400 Portable Spectrophotometer Procedures
Manual.
No Cell Holder Required
1-in. (25.4-mm) round cells/vials, including AccuVac
®
Round Cell Holder, 16-mm
COD/Test ‘N Tube™ and UniCell™ vials
Metric Rectangular Cell Holder (optional)
1-cm rectangular cells
Multi-pathlength Cell Holder
Pour-Thru Cell and multi-pathlength cells
Interface
Graphic user interface 320 x 240 pixel graphic display with touchscreen
Data Storage
1000 data points (date, time, results, sample ID, user ID), 50 user calibrations
Dimensions
32.4 x 17.8 x 18.8 cm (12.75 x 7.0 x 7.4 in.)
Weight
2.1 kg (4.6 lbs)
Keypad
5 keys for power on/off, backlight on/off, contrast down, contrast up, and battery
status
Printer
External serial
Pour-Thru Cell
1-inch/1-cm multi-pathlength (optional)
Report Generation
Download stored information in a standard report format. Meets GLP
guidelines, including a minimum of: date, time, sample identification tag, analyst
initials (3), results, and instrument serial number.
Real-time Clock
Yes
Real-time Calendar
Yes
Operating Environment
0 to 40 °C (32 to 104 °F); 90% relative humidity, non-condensing
Storage Environment
–20 to 60 °C (–4 to 140 °F); 85% relative humidity, non-condensing
Содержание DR2400
Страница 2: ...Page 2 5940018Title fm...
Страница 6: ...Page 6 5940018TOC fm...
Страница 8: ...Page 8 5940018Safety fm...
Страница 10: ...Page 10 5940018Spec fm...
Страница 12: ...Page 12 5940018InstStop fm...
Страница 20: ...Page 20 5940018Getting Started fm...
Страница 32: ...Page 32 5940018Operation fm...
Страница 46: ...Page 46 5940018StdAdd fm...
Страница 48: ...Page 48 5940018SingleWave fm...
Страница 50: ...Page 50 5940018MaintStop fm...
Страница 56: ...Page 56 5940018Troubleshoot fm...
Страница 58: ...Page 58 5940018GenInfo fm...