 Start the instrument

timer. A 15-minute reaction
time starts.

18. 

 When the timer expires,

set the heater power to off.
Remove the cylinder/gas
bubbler assembly as a unit.

19. 

 Move the gas bubbler

up and down in the arsenic
absorber solution to rinse
the bubbler.

20. 

 

Prepare the blank:

 Fill

a dry 10-mL sample cell with
unreacted arsenic absorber
solution. Put the stopper on
the sample cell.

21. 

 Clean the blank sample

cell.

22. 

 Insert the blank into the

cell holder.

Zero

23. 

 Push 

ZERO

. The

display shows the (non-
zero) intercept that is
calculated from the user-
entered calibration curve.

24. 

 

Prepare the sample:

Pour the reacted arsenic
absorber sample into a
sample cell. Put the stopper
on the sample cell.

25. 

 Clean the prepared

sample cell.

26. 

 Insert the prepared

sample into the cell holder.

Read

27. 

 Push 

READ

. Results

show in mg/L As.

Interferences

Interfering substance

Interference level

Antimony Salts

Can interfere with color development.

4

Arsenic Silver Diethyldithiocarbamate Method (0.200 mg/L)

Содержание DR 6000

Страница 1: ...Before starting A manual calibration is necessary for this procedure refer to Calibration on page 5 Make a new calibration and update the user program for each new lot of arsenic absorber solution Pre...

Страница 2: ...p the preserved samples at room temperature for a maximum of 6 months Before analysis adjust the pH to approximately 7 with 5 N sodium hydroxide solution Correct the test result for the dilution cause...

Страница 3: ...ric Acid ACS to the distillation flask 10 Use a serological pipet to add 1 mL of Stannous Chloride Solution to the distillation flask 11 Use a serological pipet to add 3 mL of Potassium Iodide Solutio...

Страница 4: ...n the blank sample cell 22 Insert the blank into the cell holder Zero 23 Push ZERO The display shows the non zero intercept that is calculated from the user entered calibration curve 24 Prepare the sa...

Страница 5: ...s follows a Use a pipet to add 10 0 mL of the 1000 mg L Arsenic Standard Solution into a 1000 mL volumetric flask b Dilute to the mark with deionized water 2 Prepare three calibration standard solutio...

Страница 6: ...FORCE ZERO is off 12 If the graph is acceptable select DONE EXIT 13 Save store the user program When the user program is selected in the test procedure the calibration curve is used to measure the sa...

Страница 7: ...s stopper 1 each 1457453 Flask volumetric Class A 200 mL 1 each 1457445 Flask volumetric Class A 500 mL 6 each 1457449 Pipet filler safety bulb 1 each 1465100 Pipet serological 5 mL 1 each 53237 Pipet...

Страница 8: ...ANCE PRICE INFORMATION AND ORDERING Call 800 227 4224 Contact the HACH office or distributor serving you www hach com techhelp hach com In the U S A Outside the U S A On the Worldwide Web E mail toll...

Отзывы: