Table 8 Navigation bar description (continued)
Icon
Description
Sets the
• General settings
• Measurement intervals
• Cleaning intervals
• Analog outputs
• Relay settings
Sets the display to off or starts the controller again.
Startup
Supply power
Connect the AC power plug to an electrical outlet to supply power to the instrument. If hard-wired
with conduit, use the switch at the local disconnect to supply power. For DC-powered instruments,
use the switch at the local disconnect to supply power.
Connect a sensor
N O T I C E
Never connect the sensor to the controller when the controller is energized or damage to the sensor can occur.
1.
Set the controller to off.
2.
Connect the sensor cable to one of the sensor connectors. Refer to
on page 6.
3.
Set the controller to on.
4.
Push
Sensor
in the navigation bar.
5.
Select the COM left or COM right button for the sensor connector. Enter the settings that follow.
Option
Setting
Operation Mode
Sensor
Hardware mode
RS-485
Protocol
Modbus
Baud rate
9600
Flow control
None
Parity
None
Stop bits
1
6.
When a sensor is connected to the controller, the sensor is shown below the corresponding COM
port. If the sensor is not seen, push
Scan
and do steps
to
7.
Select a parameter to configure the offset, scaling and warning thresholds.
20
English
Содержание CD300
Страница 2: ...English 3 Deutsch 25 Italiano 49 Fran ais 72 Espa ol 94 Portugu s 116 2...
Страница 11: ...English 11...
Страница 33: ...Deutsch 33...
Страница 57: ...Italiano 57...
Страница 80: ...80 Fran ais...
Страница 102: ...102 Espa ol...
Страница 124: ...124 Portugu s...
Страница 138: ......
Страница 139: ......