Page 95
8.1.4 Data Output
D A T A O U T P U T 0 9 : 1 7 : 2 8 1 2 - 0 9 - 0 2
1 < O U T P U T D E V I C E M M C / S D C A R D
2 S E N D R E A C T I O N A R C H I V E
3 S E N D F A U L T A R C H I V E
4 S E N D C O N F I G U R A T I O N
5 S E N D A L L D A T A
6
7 - - > D A T A P R O G R A M
This menu enables the user to select the communication port and to download the contents of the system
reaction, fault archives, system specific configuration and all data for diagnostics purposes.
1. Output Device.
This item allows the user to select communication port configuration profile. Available
option is MMC/SD CARD. See Section
for specific output device settings.
2. Send Reaction Archive.
Sub menu used to download the reaction archive to the selected output device.
3. Send Fault Archive.
Sub menu used to download the fault archive to the selected output device.
4. Send Configuration.
Sub menu used to download the system configuration to the selected output
device.
5. Send All Data.
Sub menu used to download system all data, which includes system configuration, fault
archive, reaction archive and system diagnostics information.
7.
Data Program.
Data Program is a link to Maintenance, Commissioning, Data Program menu (see
When external MMC/SD memory card is used as the output device, the data is downloaded into the card in
text format. Note that;
Any text data (reaction and fault archive, configuration and all data) can be downloaded into the card
while BioTector is running.
The card can be removed when BioTector is running.
The card should not be removed before the data transfer is completed.
If the data download into the card is successful, the files, which can be accessed in the memory card
in text format, are reaction archive, fault archive, configuration and/or all data.
Other files which are located in the in system’s external memory card by default are system firmware
(sysfrmw.hex) and system configuration (syscnfg.bin) both of which are in binary formats. Binary files
can only be opened and viewed by specific computer programs. Therefore the user should not
attempt to open or access these files.
The memory card used in BioTector can be an MMC/SD card formatted with FAT, FAT12/16 or
FAT32 file systems. Most SDHC cards are also supported and can also be used.
Содержание BioTector B3500C
Страница 17: ...Page 17 Software Menu Diagram...
Страница 44: ...Page 44 Figure 4 BioTector analysis layout typical TIC TOC system...
Страница 46: ...Page 46 Figure 6 BioTector oxygen concentrator layout...
Страница 63: ...Page 63...
Страница 78: ...Page 78 Section 8 Maintenance Menu Maintenance Menu Diagram...
Страница 155: ...Page 155 Section 11 System Replacement and Spare Parts...
Страница 163: ...Page 163 ZK Zero check ZM Manually input zero adjust ZS Zero and Span...