Table of contents
on page 3
on page 5
on page 10
Specifications
Specifications are subject to change without notice.
Table 1 General specifications
Specification
Details
Dimensions (W x D x
H)
626.11 x 527.05 x 522.73 mm (24.65 x 20.75 x
20.58 in.)
Enclosure
Rating: UL Type 4X (indoor), IEC 529 (IP 66) when air
purge is used and the door is closed and latched
Material: PC/ABS case, PC door, PC hinges and
latches, 316 SST hardware
Weight
22.4 kg (49.3 lb) dry, 28.9 kg (63.7 lb) wet with full
analyzer bottles
1
(approximately)
Mounting
Wall or bench
Protection class
I
Pollution degree
2
Installation category
II
Power requirements
100–240 VAC, 50/60 Hz, 100 VA
Connection: 18 to 14 AWG wire, 18 AWG stranded
recommended
Operating temperature
5 to 50 °C (41 to 122 °F); best performance: 5 to 35 °C
(41 to 95 °F)
Operating humidity
5 to 95% relative humidity, non-condensing
Maximum altitude
2000 m (6560 ft)
Table 1 General specifications (continued)
Specification
Details
Recorder outputs
Two 4–20 mA outputs applicable for recorders; load
impedance: 675 Ω maximum; 130 VAC isolation from
earth ground
Connection: 22 to 14 AWG wire, 22 to 20 AWG
recommended, twisted pair shielded wire
Alarm relay outputs
Two dry contact SPDT relays, not powered, each rated
at 5 A resistive, 240 VAC maximum
Connection: 18 to 14 AWG wire, 18 AWG stranded
recommended
Analyzer bottles
1 liter, high-density polyethylene with polypropylene
caps
Fuses
Two, 1.6 A, 250 V, 5 mm x 20 mm
External plumbing
connections
Sample lines: 6.35 mm (0.25-in.) OD push-in fitting for
poly tubing
Chemical drain (drain block): 12.7 mm (0.5-in.) ID barb
fitting for hose
Case drain: 25.4 mm (1-in.) ID for hose
Bottle enclosure drains: 9.53 mm (0.375-in.) ID hose
Air purge line: 6.35 mm (0.25-in.) OD push-in fitting for
poly tubing
Sample pressure
No minimum or maximum requirement
Sample flow rate
100–2000 mL/minute; 500–1000 mL/minute
recommended
Sample temperature
5 to 50 °C (41 to 122 °F)
Sample particle size
0.5 µm maximum; sample goes through a 0.5 µm filter
in the sample conditioning block before analysis
Number of sample
streams
1; optional sample sequencing kit adds a second
sample stream, sample sequence is programmable
Node-to-node distance
400 m (1320 ft) maximum between analyzers
English
3
Содержание APA 6000
Страница 1: ...DOC023 97 80365 APA 6000 05 2013 Edition 1 Installation Installation Instalaci n...
Страница 2: ...English 3 Fran ais 21 Espa ol 40 2...
Страница 30: ...Figure 9 Montage sur une table Figure 10 Montage sur un mur 30 Fran ais...
Страница 49: ...Figura 9 Montaje de sobremesa Figura 10 Montaje en pared Espa ol 49...
Страница 60: ......
Страница 61: ......