Calibrate the analog outputs
If the analog output connections of the controller are used, calibrate the
analog outputs as necessary. The analog outputs are factory-calibrated.
The adjustment range for analog output calibration is ±2 mA.
Note: If an analog output is configured to be 0–20 mA, 4 mA and 20 mA (not the
0 mA) are calibrated.
1.
Push
MENU
and select CALIBRATE>CAL OUTPUTS>CAL OUTPUT
1 (or 2).
2.
Select a calibration option.
Option
Description
CAL OUT
1 4 mA
Set the count value that represents 4 mA on the analog
outputs—550 to 910. Use a calibrated digital multimeter to
measure the actual minimum value supplied at the analog
output connections of the controller. Adjust the count value
until the analog outputs are 4.00 mA.
CAL OUT
1 20 mA
Set the count value that represents 20 mA on the analog
outputs—3460 to 3820. Use a calibrated digital multimeter to
measure the actual maximum value supplied at the analog
output connections of the controller. Adjust the count value
until the analog outputs are 20.00 mA.
3.
Push
ENTER
to confirm.
Reset the calibration
To set the calibration to the factory defaults:
1.
From the Measure screen, push
MENU
and select
TEST/MAINT>RESET CAL.
2.
Push
ENTER
to confirm.
Note: Push ESC to not set the calibration to the factory defaults.
Enable or disable the passcode
Enable the passcode to prevent unauthorized changes to the
configuration settings and calibrations. The passcode is disabled by
default. The passcode is 3456 and cannot be changed.
1.
Push
MENU
and select CONFIGURE>SET PASSCODE.
2.
Push the
UP
or
DOWN
arrow to select the option.
Option
Description
Disabled
The passcode does not have to be entered to change the
configuration settings, do calibrations or go to the TEST/MAINT
screens.
Enabled
The configuration settings can be seen but not changed until the
passcode is entered. The CALIBRATE and TEST/MAINT screens
cannot be seen until the passcode is entered.
3.
Push
ENTER
to confirm.
Maintenance
W A R N I N G
Multiple hazards. Only qualified personnel must conduct the tasks
described in this section of the document.
Clean the instrument
Clean the exterior of the instrument with a moist cloth and a mild soap
solution and then wipe the instrument dry.
Clean the flow chamber
Clean the flow chamber to remove unwanted material (e.g., dirt, sand)
from the flow chamber.
1.
Drain the flow chamber.
a.
Turn the internal 3-way valve handle to the CALIBRATE position.
b.
Remove the funnel from the holder and hold it below the sensor
height to drain any sample that is still in the sensor.
c.
When the sensor is drained, put the funnel in the funnel holder.
2.
Remove the flow chamber cover. Turn the cover
1
/
8
turn counter-
clockwise to unlock it, then lift the cover off.
English
23
Содержание Accu4
Страница 1: ...DOC023 97 80356 Accu4 T53 8320 01 2013 Edition 1 User Manual...
Страница 2: ...English 3 29 2...
Страница 32: ...1 8320 1 6 2 2 7 3 2 8 in NPT 4 9 5 in NPT 10 2 3 2 1 3 4 2 4 32...
Страница 34: ...4 NEMA IP NEMA 4X IP66 3H1091 3H1230 4 3 11 4 mm 0 17 0 45 in ESD ESD 5 34...
Страница 35: ...5 1 2 5 2 RS232 6 2 3 TTL 7 4 4 8 6 NEMA IP 6 35...
Страница 37: ...7 2 TB 1 12 18 13 19 14 20 15 21 16 22 0 00 20 00 mA 4 00 20 00 mA 1 2 5 35 35 42 600 37...
Страница 40: ...10 1 in NPT 4 in IPS 2 in IPS 5 3 6 in NPT 11 11 1 2 12 T53 4 40...
Страница 50: ......
Страница 51: ......