Calibrate

Items to collect:

 One auto-recognition calibration standard

1.

Set the power to on.

2.

Remove the cap from the sensor.

3.

Push   to go to calibration mode.
The auto-recognition standard(s) the tester expects to measure shows on the bottom line.

4.

Rinse the sensor and cap with deionized water and blot dry.

5.

Pour the calibration standard shown into the cap to the fill line.

6.

Put the sensor fully into the cap.

7.

When the measurement is stable, push   to save the calibration and go to continuous measurement mode.
The measured value will flash 3 times and then stop. Then, "END" shows on the display.

8.

Rinse the sensor and cap with deionized water and blot dry.

Measurement

Note: Air bubbles under the probe tip when submerged can cause slow stabilization or error in measurement. Shake the tester

from side to side to remove air bubbles.

1.

Set the power to on.

2.

Remove the cap from the sensor.

3.

If the lock icon shows on the display, push   to go to continuous measurement mode.

4.

Rinse the sensor and cap with deionized water and blot dry.

5.

Pour the water sample into the cap to the fill line.

6.

Put the sensor fully into the cap. The measured value shows on the top line.

7.

To keep the measured value on the display when the sensor is removed from the sample, push  .

Note: The lock icon shows on the display when the measurement is stable.

8.

To measure another sample, do steps 

3

7

.

9.

When done with measurements:

a.

Rinse the sensor and cap with deionized water.

b.

Put the cap on the tester.

c.

Set the power to off.

Advanced operation

Configure the settings

1.

Push and hold   until "SEt" shows on the display.

2.

Push   to scroll through the settings. The current value of the setting shows on the bottom line.

Option Description

Unit

Select the temperature unit that shows on the display— C (Celsius) or F (Fahrenheit).

Unit

Select the salinity unit that shows on the display— ppt or %.

AOFF

Set the auto-off feature to On (enable, default) or Off (disabled). When set to On, power is automatically set to
off after 8 minutes of no activity.

rSEt

Change the settings to the factory defaults—Yes or No (default). When set to Yes, changes the settings to the
factory settings and default values.

3.

To change the value of the setting, push 

 or 

.

4.

When done with changes, push and hold   until "End" shows to go to continuous measurement mode.

7

Содержание 9531600

Страница 1: ...User Manual DOC022 53 80400 Pocket Pro Salinity 1...

Страница 2: ...s it describes at any time without notice or obligation Revised editions are found on the manufacturer s website Safety information The manufacturer is not responsible for any damages due to misapplic...

Страница 3: ...e complies with Part 15 of the FCC Rules Operation is subject to the following conditions 1 The equipment may not cause harmful interference 2 The equipment must accept any interference received inclu...

Страница 4: ...ged contact the manufacturer or a sales representative immediately Figure 2 Product components 1 Pocket Pro tester 3 Lanyard 2 AAA alkaline batteries 4x Install the batteries CAU TI ON Explosion hazar...

Страница 5: ...erface and navigation Display description Figure 4 shows the measurements calibration standard information and indicator icons shown on the display Table 1 gives descriptions of the icons Figure 4 Dis...

Страница 6: ...on the display to go to the settings menu To exit the settings menu push and hold until End shows on the display When in the settings menu push to scroll through the settings Note Power cannot be set...

Страница 7: ...ater sample into the cap to the fill line 6 Put the sensor fully into the cap The measured value shows on the top line 7 To keep the measured value on the display when the sensor is removed from the s...

Страница 8: ...e measurement range of the tester Gently clean the sensor Refer to Clean the sensor on page 8 Calibrate as necessary bottom line The temperature value is not within the operating temperature range of...

Страница 9: ...Table 4 Replacement parts Description Quantity Item no AAA alkaline batteries 4 pkg 4674300 Lanyard 1 201305 SINGLET 12 88 mS cm 20 pkg 2771520 9...

Страница 10: ...hach com HACH LANGE GMBH Willst tterstra e 11 D 40549 D sseldorf Germany Tel 49 0 2 11 52 88 320 Fax 49 0 2 11 52 88 210 info de hach com www de hach com HACH LANGE S rl 6 route de Compois 1222 V sen...

Отзывы: