D.10 Query
The Modbus ASCII query must take the form shown below that specifies the starting
register and number of registers to be read:
For example, to read the level channel of the 940 Flow Meter, the query must be as
indicated in
.
The master queries the flow meter using a Read Holding Registers request, which implies
a 4XXXX register reference, to slave device address 01. The message requests data from
holding registers 40007–40008 to obtain the level information, which requires two
registers to store the floating point value. Note that registers are referenced from zero in
the data field.
CFM 17
DEGREE_F
42
CFH 18
MILS 43
CFD 19
VOLTS
44
CMS 20
FPS 45
CMM 21
MPS 46
CMH 22
PCT_O2
47
CMD 23
PCT_H2S
48
GPS 24
PCT_LEL
49
GPM 25
VDC 50
Table 17 SCADA-Modbus Units of Measure Codes
Unit
Code
Unit
Code
Start
‘:’
Slave
Address
Function
(03)
Start
Address
High
Start
Address
Low
No. of Pts.
High
No. of.
Pts. Low
LRC
<CR>
<LF>
Table 18 Channel Query to Read Level (Example)
Start
‘:’
Slave Address
01
Function
03
Starting Address High
00
Starting Address Low
06
No. of Registers High
00
No. of Registers Low
02
LRC
F4
Stop
<CR><LF>
Содержание 911
Страница 2: ......
Страница 14: ...12 General Information...
Страница 16: ...14 Getting Started...
Страница 42: ...40 Meter Installation...
Страница 44: ...42 Software and Communications...
Страница 58: ...56 Sensor Installation...
Страница 68: ...66 Maintenance...
Страница 72: ...70 Replacement Parts and Accessories...
Страница 76: ...74...
Страница 88: ...86 Channel Installation Options...
Страница 92: ...90 Batteries and Chargers...