25
Section 5
Operation
5.1 Sensor Setup
When a sensor is initially installed, the serial number of the Digital Gateway will be
displayed as the sensor name. To change the sensor name refer to the following
instructions:
1.
Select Main Menu.
2.
From the Main Menu, select SENSOR SETUP and confirm.
3.
Highlight the appropriate sensor if more than one sensor is attached and confirm.
4.
Select CONFIGURE and confirm.
5.
Select EDIT NAME and edit the name. Confirm or cancel to return to the Sensor
Setup menu.
5.2 Sensor Data Logging
The sc1000 controller provides a data log for each sensor.
The sc100 provides three data logs (one for each sensor and one for calculated values).
The data logs store the measurement data at selected intervals. The event log stores a
variety of events that occur on the devices such as configuration changes, alarms, and
warning conditions. The data logs are stored in a packed binary format and the event logs
are stored in a CSV format. The logs can be downloaded through the digital network port,
service port, or the IrDA port. DataCom is needed for downloading logs to a computer.
5.3
Sensor Diagnostics Menu for pH and ORP
SELECT SENSOR (if more than one sensor is attached)
STATUS
ERROR LIST
See
section 7.1 on page 35
.
WARNING LIST
See
section 7.2 on page 35
.
5.4
pH Sensor Setup Menu
SELECT SENSOR (if more than one sensor is attached)
CALIBRATE (Main Menu Item)
1 POINT AUTO
Calibration with a single buffer — normally pH 7.
2 POINT AUTO
Calibration with two buffers — normally pH 7 and pH 4 or 10.
1 POINT MANUAL
Calibration against a single known sample.
2 POINT MANUAL
Calibration against two samples, both with a known pH.
TEMP ADJUST
Adjust the displayed temperature by up to ± 15 °C.
DEFAULT SETUP
Restores the system to the original factory calibration.
CONFIGURE
EDIT NAME
Enter up to a 10-digit name in any combination of symbols and alpha or numeric characters.
SELECT MEASURE
Select the appropriate measurement units to display.
DISPLAY FORMAT
Select the measurement resolution (xx.xx pH or xx.x pH).
TEMP UNITS
Choose from the displayed options (°C or °F).
LOG SETUP
Choose SENSOR INTERVAL to set the sensor log interval or select TEMP INTERVAL to set the
temperature log interval.
Содержание 6120118
Страница 2: ...Visit us at www hach com...
Страница 6: ...Visit us at www hach com...
Страница 10: ...Visit us at www hach com...
Страница 18: ...Visit us at www hach com...
Страница 32: ...Visit us at www hach com...
Страница 36: ...Visit us at www hach com...
Страница 38: ...Visit us at www hach com...
Страница 42: ...Visit us at www hach com...
Страница 50: ...Visit us at www hach com...
Страница 52: ...Visit us at www hach com...